8 Principles of Operation
The Oximeter adopts non-invasive double wavelength to measure
SpO
2
and PR. It can perform spot and continuous measurement for
a short time.
The system consists of Central Processing Unit, Signal Collection,
Signal Input, Data Output, Display and User Input module, shown
as follows:
Figure 8-1 System Principle
The oximeter communicates with external devices through
RS-232 interface.
8.1 Pulse Oximetry Measurement
The oximeter uses oximetry to measure functional oxygen
saturation in the blood. Pulse oximetry works by applying the
sensor to a pulsating arteriolar vascular bed, such as a finger or a
toe. The sensor contains a dual light source and a photonic
detector.
Bone, tissue, pigmentation, and venous vessels normally absorb a
constant amount of light over time. The arteriolar bed normally
pulsates and absorbs variable amounts of light during the
pulsations. The ratio of light absorbed is translated into a
measurement of functional oxygen saturation (SpO
2
). Because a
Содержание HS-Vet
Страница 1: ......
Страница 65: ...64 Appendix Ⅲ Record Table ID No Name Time SpO2 PR NOTE ...
Страница 98: ...97 Configuración del sistema Configuración de alarma ...
Страница 111: ...110 lectura manual debe corresponder con la lectura de la frecuencia del pulso que aparece en el monitor ...
Страница 119: ...118 Use un adhesivo para el sensor Use un nuevo sensor con adhesivo fresco Mantenga al animal inmóvil si es posible ...
Страница 136: ...135 pulsioxímetro ...
Страница 139: ...138 Apéndice III Tabla de registro Nº ID Nombre Hora SpO2 PR NOTA ...
Страница 204: ...203 Annexe Ⅲ Record Table No ID Nom Temps SpO2 PR REMARQUE ...
Страница 237: ...236 Alarm Setup ...
Страница 278: ...277 Anhang III Tabelle für Aufzeichnungen ID Nr Bezeichnung Zeit SpO2 PF HINWEIS ...