42
Unconnected
SpO
2
Sensor off
Low Battery
Alarm Indicators
When an alarm occurs, the oximeter will indicate it through the
following indications:
■
Character flash
■
Alarm tone
High level alarms: character flashes quickly and sounds
double beep;
Medium level alarms: character flashes slowly and sounds triple
beep;
The sound pressure range for auditory alarm signal is from 45dB
to 85dB.
5.2 Alarm Conditions
5.2.1 Alarm off Before the First Measurement
Before the first measurement, the alarm system is configured to be
off. At this time, if the SpO
2
sensor is unconnected or the sensor is
off, the oximeter will not give an alarm.
5.2.2 Alarm for SpO2 Sensor Unconnected
When the SpO
2
sensor is disconnected, the oximeter gives a
Medium alarm. The icon
displays in information area.
SpO
2
and PR value area display “---”, and give a medium alarm.
(Make sure the alarm system in menu is ON.)
5.2.3 Alarm for SpO2 Sensor off
When the SpO
2
sensor falls off from the finger, the oximeter will
give a medium alarm, and the icon
displays in
information area.
SpO
2
and PR value area display “---”, and give a medium alarm.
(Make sure the alarm system in menu is ON.)
5.2.4 Alarm for Low Battery
Содержание HS-Vet
Страница 1: ......
Страница 65: ...64 Appendix Ⅲ Record Table ID No Name Time SpO2 PR NOTE ...
Страница 98: ...97 Configuración del sistema Configuración de alarma ...
Страница 111: ...110 lectura manual debe corresponder con la lectura de la frecuencia del pulso que aparece en el monitor ...
Страница 119: ...118 Use un adhesivo para el sensor Use un nuevo sensor con adhesivo fresco Mantenga al animal inmóvil si es posible ...
Страница 136: ...135 pulsioxímetro ...
Страница 139: ...138 Apéndice III Tabla de registro Nº ID Nombre Hora SpO2 PR NOTA ...
Страница 204: ...203 Annexe Ⅲ Record Table No ID Nom Temps SpO2 PR REMARQUE ...
Страница 237: ...236 Alarm Setup ...
Страница 278: ...277 Anhang III Tabelle für Aufzeichnungen ID Nr Bezeichnung Zeit SpO2 PF HINWEIS ...