32
The oximeter implements variable pulse tone and its frequency
varies with the saturation
3.3.5 Audio Paused (s)
Set the pause period for audio alarm to 30, 60, 90 or 120 seconds.
When
Alarm System
is set to
ON
, pressing the
Alarm Silence
button can turn off the audio alarm, the pause period is set by the
Audio Paused (s)
.
3.3.6 User Maintain
Enter the
User Maintain
menu by inputting “819”.
Figure 3-7 Enter Password
If the password is wrong, the following dialog box will pop up:
Figure 3-8 Wrong Password
If the password is right, the following menu will display:
Figure 3-9 User Maintain
●
Language: the user can select language to be displayed.