41
5 Alarm
5.1 Alarm Categories and Levels
Alarm Categories
The oximeter’s alarms can be classified into two categories:
physiological alarms and technical alarms.
1.
Physiological alarms
Physiological alarms, also called patient status alarms, are
triggered by a monitored parameter value that violates setup
alarm limits or an abnormal patient condition.
2.
Technical alarms
Technical alarms, also called system status alarms, are
triggered by a device malfunction or a patient data distortion
due to improper operation or system problems.
Alarm Levels
In terms of severity, the oximeter’s alarms levels can be classified
into two categories: high level alarms and medium level alarms.
1.
High level alarms
Indicate that the patient is in a life threatening situation and
an emergency treatment is demanded.
2.
Medium level alarms
The patient’s vital signs appear abnormal or the oximeter
system status appears abnormal, indicate that prompt operator
response is required.
The levels for both technical alarms and physiological alarms are
predefined and can not be changed by the user.
Alarm Categories Table
High level alarms
Medium Level
Alarms
Physiological alarms
SpO
2
Too High
SpO
2
Too Low
PR Too High
PR Too Low
Technical alarms
SpO
2
Sensor
Содержание HS-Vet
Страница 1: ......
Страница 65: ...64 Appendix Ⅲ Record Table ID No Name Time SpO2 PR NOTE ...
Страница 98: ...97 Configuración del sistema Configuración de alarma ...
Страница 111: ...110 lectura manual debe corresponder con la lectura de la frecuencia del pulso que aparece en el monitor ...
Страница 119: ...118 Use un adhesivo para el sensor Use un nuevo sensor con adhesivo fresco Mantenga al animal inmóvil si es posible ...
Страница 136: ...135 pulsioxímetro ...
Страница 139: ...138 Apéndice III Tabla de registro Nº ID Nombre Hora SpO2 PR NOTA ...
Страница 204: ...203 Annexe Ⅲ Record Table No ID Nom Temps SpO2 PR REMARQUE ...
Страница 237: ...236 Alarm Setup ...
Страница 278: ...277 Anhang III Tabelle für Aufzeichnungen ID Nr Bezeichnung Zeit SpO2 PF HINWEIS ...