23
Checking the Ni-MH Battery Package
The performance of a Ni-MH rechargeable battery package may
deteriorate. To check the performance of the battery, follow the
procedures below:
1.
Disconnect the pulse oximeter from the patient and stop all
monitoring and measuring procedures.
2.
Place the pulse oximeter in the charger stand and connect the
AC mains. Allow the battery to be charged uninterruptedly
for above 2.5 hours. For details about charging the Ni-MH
battery package, please refer to section 3.4.
3.
Disconnect AC mains and allow the pulse oximeter to run in
the measurement state until it shuts off.
The operating time of a battery reflects its performance directly. If
the operating time of a Ni-MH rechargeable battery package is
noticeably shorter than that stated in the specifications, replace it
or contact your service personnel.
Low Battery Icon
The low battery icon displays and an alarm is given when
few-minute operation remains available. After few-minute
operation, the oximeter will turn off automatically. Replace the
Содержание HS-Vet
Страница 1: ......
Страница 65: ...64 Appendix Ⅲ Record Table ID No Name Time SpO2 PR NOTE ...
Страница 98: ...97 Configuración del sistema Configuración de alarma ...
Страница 111: ...110 lectura manual debe corresponder con la lectura de la frecuencia del pulso que aparece en el monitor ...
Страница 119: ...118 Use un adhesivo para el sensor Use un nuevo sensor con adhesivo fresco Mantenga al animal inmóvil si es posible ...
Страница 136: ...135 pulsioxímetro ...
Страница 139: ...138 Apéndice III Tabla de registro Nº ID Nombre Hora SpO2 PR NOTA ...
Страница 204: ...203 Annexe Ⅲ Record Table No ID Nom Temps SpO2 PR REMARQUE ...
Страница 237: ...236 Alarm Setup ...
Страница 278: ...277 Anhang III Tabelle für Aufzeichnungen ID Nr Bezeichnung Zeit SpO2 PF HINWEIS ...