29
Time scale Activity
Daily
Clean the vacuum chamber, lid, and housing after use with a soft
damp cloth.
Make sure that no cleaning agents containing solvents are used.
Make sure that no high pressure water cleaner is used.
Weekly
Check the oil level and replace or fill up oil when the oil is turbid or the
oil level is too low. See instructions.
Activate the pump cleaning program for the pump.
Inspect the sealing bar for damage. Replace Teflon tape/sealing wire
if the seal quality is no longer sufficient or if the Teflon tape/sealing
wire is no longer tight and straight on the sealing bar. See instructions.
Inspect the lid gasket and replace it when the gasket is damaged. See
instructions.
Inspect the transparent lid. When cracks are visible, turn off the
machine immediately and contact the supplier.
Every Six
Months
Replace oil.
Replace rubber in the silicone holder.
Yearly
Inspect the oil exhaust filter for saturation. If saturated, replace the
filter. See instructions.
Contact the supplier for a professional service.
Four-yearly
Replace transparent lid and the lid’s gas springs.
Replace membranes seal cylinder.