37
WE2108
HBM
A1632-2.0 en
”Calibration” submenu (see Section 6 ”Calibration”)
(only with access authorization!)
Values can be changes manually or automatically measured with
ENTER
ENTER
CE
NEXT
UP
Para-
meter
Meaning
Value range
Remarks
Lin
Linearization
0: off
(2 Points)
1: polynominal (3 points)
2: polynominal (4 points)
0: Part load calibration
possible
CAL−0
Internal value, unloaded
−20000...20000 (=2mV/V)
Optional input or
automatic measurement
LOAd1
Display for weight 1
0 ... 99999
e.g. half nominal value in
3−point calibration
CAL−1
Internal value for weight 1
−20000...20000 (=2mV/V)
Optional input or
automatic measurement
LOAd2
Display for weight 2
0 ... 99999
Blocked for ”Lin” = 0
CAL−2
Internal value for weight 2
−20000...20000 (=2mV/V)
Blocked for ”Lin” = 0
LOAd3
Display for weight 3
0 ... 99999
Blocked for ”Lin” = 0 or 1
CAL−3
Internal value for weight 3
−20000...20000 (=2mV/V)
Blocked for ”Lin” = 0 or 1
Factory setting
ENTER
CE
Para-
meter
Meaning
Value range
Remarks
dEFLt
Resetting the settings (e.g.
filter) to factory default
−−
The scale calibration is de-
leted i.e. the WE2108 is set
to the factory characteristic
0 ... 2 mV/V.
Further parameters are only for service purposes
Printing Parameters
Value range
Remarks
Printing the parameters with
ENTER
−−
Appears only if the interface
is set to printing (”F_SEr” = 1
in the
ioSEt
menu)