39
WE2108
HBM
A1632-2.0 en
Summing functions
Para-
meter
Meaning
Remarks
F_Add
Switching the summing
function on/off
total
Indication of sum, if nec. in
two display phases for val-
ues > 5 digits
t_nbr
Indication of summand
counter
clr_t
Deletion of sum and
summand counter
Prtno
Indication of running num-
ber for print output
clr_P
Deletion of running num-
ber
Display lightning ON / OFF
Switch light on / off with
ENTER
−−
see 5.1.7 ”Display ligth-
ning”
Switch unit off with
ENTER
−−
see 5.1.1 ”Switching on
and off”
5.3
Explanations for the settings
5.3.1 Display
limits
The permissible range of the display depends upon the scale standard. The
setting is made in the
Func
submenu, ”trAdE” parameter.
trAdE
Scale standard
lower limit
upper limit
0
not legal for trade
−135 % nominal value
135 % of nominal value
1
OIML
−2 % nominal value
nominal value + 9d
2
NTEP
−2 % nominal value
105 % of nominal value
5.3.2 Tare
value
A switchover in the net display by means of the
G/N
−key is only possible
for a tare value (or manual tare value) > 0.
The operation of the tare key always triggers a taring of the applied load, that
is, subsequently the tare value equals the current gross value. By means of
the setting ”F_tAr” = 1 (menu
Func
, parameter−protected) a manual tare func-
tion can be activated additionally. When the
G/N
−key is operated, the man-
ual tare value, which was input in the menu ”P_tAr” , is used instead of the
tared value.