35
WE2108
HBM
A1632-2.0 en
”Filter settings” submenu
ENTER
CE
NEXT
UP
Para-
meter
Meaning
Value range
Remarks
F_FiL
Digital filter setting
(3dB – cutoff frequency
at ”Icr” = 0)
0: 13,0Hz
1: 12,5Hz
2: 11,2Hz
3: 07,5Hz
4: 03,0Hz
5: 02,5Hz
6: 02,0Hz
7: 01,0Hz
8: Special filter for
animal scales
Mesurement /
Settling time
rate [s
−1
]
[ms]
0.
50
/
090
1.
50
/
100
2.
50
/
110
3.
50
/
120
4.
50
/
400
5.
40
/
475
6.
33
/
550
7.
28
/
625
8. not specif.
/ Signal
(variable)
dependant
Icr
Mean value formation
0 ... 99
The averaged values are
output with unchanged mea-
suring rate (moving average
filter).