HB-TP180/200
Checking Facility for Temperature Control Units
Testing
38
O8354-EN 2018-11
6.
Connect the reference measuring device with the instrument
leads to the
Reference Pt 100
port.
7.
Set the following parameters as indicated.
only for Thermo-5:
Open the
Setting \ Controller
menu page.
Set the
Measuring point internal
parameter to “Main line“.
Set the
Operating mode
parameter to “manual”.
Open the
Settings \ Miscellaneous
menu page.
Set the
Pressure relief with unit OFF
parameter to “OFF”.
only for Series 4:
Open the
Settings \ Miscellaneous
menu page.
Set the
Measuring point internal
parameter to “Main line“.
Open the Setting \ Service \ Parameter \ Controller menu page.
Set the
Operating mode
parameter to “manual”.
Open the Setting \ Service \ Parameter \ System menu page.
Set the
Pressure relief with unit OFF
parameter to “OFF”.