72
an alternative you can spray the fluid onto the tissue first, and then apply it to the
filter as you wipe it.
2. Fold the tissue several times to match the width of the IR filter—you might use two
or three tissues at a time if necessary (to give you a better grip). Be sure to fold the
paper so that the coated glossy side faces outwards—do not use the other side, as
it can do more harm than good!
3. Gently place the folded tissue onto the edge of the filter using two or three fingers.
Be sure to wipe the entire surface evenly. Sweep the filter only once. Do not wipe
the same area twice with the same tissue as you might reapply dirt removed in the
first sweep.
4. Finally check if the IR filter has been properly cleaned either by visual inspection or
by mounting the digital back to the camera and making a shot. If further cleaning
is needed, repeat cleaning procedure.
Cleaning with an E-Wipe
E-wipes are individually packed wet tissues.
1. Tear at the notch to break seal.
2. Remove e-wipe from its packaging and continue without delay.
3. Fold the tissue to match the width of the IR filter.
4. Apply firm pressure using two or three fingers at the edge of the wipe to ensure an
even, firm contact with filter surface. Wipe the surface in one unbroken motion.
Note! Do not use same side of the e-wipe twice as you will be likely to reapply any particles
removed in the first pass.
5. Finally check if the IR filter has been properly cleaned either by visual inspection or
by mounting the digital back to the camera and making a shot. If further cleaning
is needed, repeat cleaning procedure.
Cleaning the housings
If the camera becomes dirty, clean it with a soft, clean cloth lightly moistened with
water only. Do not use any other solvents and do not allow water to get.