Hardy 1734-WS Installation & Operation Manual
Page | 37
cell.
Command Status 8 Hard Cal Fail
Counts
Too few A/D counts between zero
and span points during hard
calibration. Input must change by a
minimum amount between the cal
low and cal high points.
Add weight to the scale and see if
readings increase. Check voltages
as in ADC Convert Error.
Command Status 0x0B (11 dec)
Value being set is too high
Verify the value is within the
tolerance for the parameter being
written to. Verify the parameter ID
is the correct ID for the parameter
you wish to write to.
Command Status 0x0C (12 dec)
Value being set is too low
Verify the value is within the
tolerance for the parameter being
written to. Verify the parameter ID
is the correct ID for the parameter
you wish to write to.
Command Status 0x0D (13dec)
Value not allowed