Hardy 1734-WS Installation & Operation Manual
Page | 26
11 = Fail - value too high
12 = Fail – Value too low
13 = Fail – not allowed
128 = Fail – Parameter ID not found
Output Table
Output Table for the AOP:
OUTPUT table
Type
Command INT
Aux Command Information
INT
Parameter Value
DINT
Parameter ID
INT
Reserved word 1
INT
Reserved word 2
DINT
Reserved word 3
DINT
Parameter RD1 ID
INT
Reserved word 5
INT
Parameter RD2 ID
INT
Reserved word 6
INT
Total
14
Command, Aux Command Information, Parameter ID, and Parameter Value, are used to send commands to
the instrument, write new parameter values, read existing parameter values, or read data values. The
Command is a 16-bit value used for the command string as shown above in the command section.
The 16-bit Aux Command Information is used for specific information required for special commands. To
select which parameter is being read or written, set the predefined number into the Parameter ID. If the
value is being read, then the Parameter Value is ignored, or set to the required value if the value is being
written.
The following values, Reserved 1, Reserved 2, Reserved 3, Reserved 4, Reserved 5, Reserved 6, and Reserved
7, are reserved, and also provide padding so the user selectable read only parameters are aligned between
the output and input tables. The other two values Parameter RD1 ID and Parameter RD2 ID are user
selectable parameter ID values which are used to read values from the instrument. These read values can be
anything from an instrument specific measurement such as Num Averages to a parameter value such as
Waversaver.
Input Table
Input Table for the AOP:
INPUT table
Type
Command Echo
INT
Command Status
INT
Parameter Value
DINT
Parameter ID
INT
Instrument Status
INT
Net Weight
REAL
Gross Weight
REAL