background image

37

Subject to change without notice

mode is automatically set to the value used during “sea”
(search) operation.

Photo 3

MODE: “del” (DELAY)

TIME/DIV: 5ms/div

Trigger coupling: TvF

Trigger slope: falling (-)

Delay time:

4div x 5ms/div = 20ms

Reducing the time coefficient (increasing the time base speed)
now expands the signal. If the signal start position is not set to
the optimum, it can still be shifted in the

 

X

 

direction by

changing the delay time.

Photo 4 shows a 50 fold X magnification caused by setting the
time coefficient to 0.1ms/div (5ms/div : 0.1ms/div = 50). The
reading accuracy also increases with higher X magnification.

Photo 4

MODE: “del” (DELAY)

TIME/DIV: 0.1ms/div

Trigger coupling: TvF

Trigger slope: falling (-)

Delay time: 20ms

The delayed and expanded signal display can be triggered again if
a signal slope suitable for triggering appears after the delay time.
For this, one must switch to     

“dTr”

     (2nd triggering after the expiry

of the delay time - after Delay Triggering). The settings selected
before switching, automatic Peak value triggering / Normal
triggering, trigger coupling, the trigger     

LEVEL

     setting and slope

setting, remain valid and trigger the start of the delay time.

The     

”After Delay”

     Triggering automatically switches to normal

triggering (indicated by the     

NM LED

) and DC trigger coupling. These

default conditions cannot be changed. But the trigger level (

LEVEL

)

and the trigger slope direction can be altered in order to enable the
triggering at the desired signal section. The trace does not start and
the screen remains blank if the signal amplitude is not sufficient for
triggering or if the setting of the trigger     

LEVEL

     is unsuitable.

The expanded display can also be displaced in the X direction by
changing the delay time (

DEL/TR. POS.

) under suitable settings.

However, the displacement is not continuous as in the untriggered

“del” (DELAY)

     operation but jumps from one trigger slope to another

- with most signals this is not evident. This means, in the case of  TV
Triggering, that it is possible to trigger not only with line synchronizing
pulses but also on suitable slopes occurring within the line.

Of course, the magnification is not restricted to a factor 50 as
mentioned in the example. The limit is given by the increasing
loss of trace intensity as the magnification is increased.

The manipulation of time delay requires a certain experience,
especially with  complicated signal combinations which are
difficult to display. The display of sections of simple signals  is,
in contrast, fairly easy. The time delayed display is also possible
in the dual channel, addition and difference modes.

In chopped DUAL mode, if after switching over to “del” or “dTr”,
the time deflection coefficient is reduced (TIME/DIV.), the channel
switching mode doesn`t change automatically to alternate.

Photo 1

(composite video signal)

MODE: “DEL. MODE” OFF

TIME/DIV: 5ms/div

Trigger coupling: TvF

Trigger slope: falling (-

)

Switching over from undelayed to delayed time base     automatically
sets the hold off time to minimum so that the HO LED extinguishes,
the DEL/TR. POS. knob function changes from hold off time to
delay time control and the READOUT indicates “sea”.

In search (“sea”) mode a part of the previously complete visible
trace becomes blank. The length of the blanked sector depends on
the delay time     

(DEL/TR. POS.)

     setting and can be set between

approx. two and seven divisions after the normal trace start position.
Consequently the trace is displayed with reduced length.

If the maximum delay time is not sufficient, the time coefficient
must be increased (

TIME/DIV

 knob) and the     

DEL/TR. POS.

knob set to the later starting point.

Note:
Actually the trace start is not really delayed in “sea” (search)
condition, as the blank sector serves only as an adjusting
indicator making visible the delay time which will be active
after selecting “del” (delay time base) or “dTr” (delay time
base in triggered condition).

Photo 2

MODE: “sea” (SEARCH)

TIME/DIV: 5ms/div

Trigger coupling: TvF

Trigger slope: falling (-)

Delay time:

4div x 5ms = 20ms

Figure 2 shows that the delay time can be measured. It is
identical with the displacement of the start of the trace. One
can calculate this by multiplying the blanked out section
(horizontal) by the time deflection coefficient setting.

The full length trace will be visible when switched from “sea”
(SEARCH) to “del” (DELAY), starting with the section previously
selected, providing the (stored) current time deflection
coefficient is not too small.

If the trace is invisible or hardly visible because of too much expansion
(too small deflection coefficient), the time deflection coefficient must
be increased with     

TIME / DIV

     knob. A larger deflection coefficient

than in the     

“sea” (SEARCH) 

mode cannot be set.

Example:
The SEARCH setting selected in figure 2  is 5ms/cm.
The display in “del” (DELAY) mode, also with 5ms/div
is delayed but unexpanded (1:1). A further increase in
the deflection coefficient, e.g. 10ms/div would be
meaningless and therefore automatically blocked.

Please note that the previous time coefficient chosen in
“del”

“del”

“del”

“del”

“del”  and 

“dTr”

     mode is stored and automatically set after

activating one of those modes. If the stored time coefficient
in  “del” / 

  “del” / 

  “del” / 

  “del” / 

  “del” / 

”dTr”

     mode was higher than the actual value in

“sea” 

“sea” 

“sea” 

“sea” 

“sea” (search) mode, the time coefficient in     

“del” / ”dTr”

Triggering and time base

Содержание HM507

Страница 1: ...Instruments Oscilloscope HM507 E N G L I S H HANDBUCH MANUAL MANUEL...

Страница 2: ...MANUAL HANDBUCH MANUEL...

Страница 3: ...tomatic Peak value Triggering 33 Normal Triggering 34 Slope 34 Trigger coupling 34 Triggering of video signals 34 Line Mains triggering 35 Alternate triggering 35 External triggering 35 Trigger indica...

Страница 4: ...place via the device under test mains line supply test leads control cables and or radiation The device under test as well as the oscilloscope may be effected by such fields Although the interior of t...

Страница 5: ...Interface 50MHz Analog Digital Oscilloscope HM507 Oscilloscope Specifications Vertical Deflection analog digital Operating modes Channel I or CH II separate CH I and II alternate or chopped 0 5MHz Sum...

Страница 6: ...be negated by the use of an extension cord without a protective conductor The mains line plug should be inserted before connections are made to measuring circuits The grounded accessible metal parts c...

Страница 7: ...data are based Purchase of the new HAMEG scope tester HZ 60 which despite its low price is highly suitable for tasks of this type is very much recommended The exterior of the oscilloscope should be cl...

Страница 8: ...lue is 11 higher The gain reduction error can not be defined exactly as the 3dB bandwidth of the amplifiers differ between 40MHz and 42MHz For sinewave signals the 6dB limit is approx 50MHz When exami...

Страница 9: ...se four values are not freely selectable They have to be within the following limits L between 0 2 and 10div if possible 4 to 10div T between 10ns and 5s F between 0 5Hz and 100MHz Tc between 100ns di...

Страница 10: ...ng the probe should always be used The series impedance of the probe provides a certain amount of protection for the input of the vertical amplifier Because of their separate manufacture all attenuato...

Страница 11: ...at 1 Component Tester is switched off 2 The following settings are present under MAIN MENU SETUP INFO MISCELLANEOUS 2 1 CONTROL BEEP and ERROR BEEP activated x 2 2 QUICK START not activated 3 The scre...

Страница 12: ...memories These can be used by the operator to save instrument settings and to recall them SAVE Press and hold the RECALL SAVE button to start a storage process This causes the SAVE menu Standard menu...

Страница 13: ...e determined by taking the deflection coefficient the probe factor and the trace position change with respect to the previous 0 Volt position into account 0 Volt Symbol The READOUT indicates the 0 Vol...

Страница 14: ...tomatic or automatic peak value triggering is selected AT Automatic triggering can be carried out with or without peak capture In both cases the LEVEL control 11 is effective and the trace is visible...

Страница 15: ...direction ccw The available range is from 1mV div up to 20V div The deflection coefficients and additional information regarding the active channel s are displayed in the readout e g Y1 deflection coe...

Страница 16: ...coefficient display as the time base is switched off 2 Controls and readout display switched off for trigger source slope and coupling 3 Y POS CURS I knob inactive for X shift use X POS 12 knob 4 X MA...

Страница 17: ...nel II and the READOUT displays Y2 trigger slope trigger coupling The last function setting of the VOLTS DIV 18 knob remains unchanged All channel II related controls are active if the input 28 is not...

Страница 18: ...zero and maximum can be set The signal displayed to the left of the trigger point symbol shows the pre history the right side displays the run of the curve after the trigger event POST TRIGGER As defi...

Страница 19: ...cient and can be selected by the TIME DIV knob It is recommended to capture and display the signals in DUAL mode and choose a suitable time coefficient which enables you to see the higher frequency si...

Страница 20: ...7 divisions The blanked section serves as a guide to the delay time The delay time is based on the current time deflection coefficient setting and can also be coarsely set with the TIME DIV control ra...

Страница 21: ...in conjunction with an operating mode instead of a measuring value the READOUT indicates n a not applicable For example the READOUT displays t n a if t measurement is selected in combination with XY m...

Страница 22: ...er and trigger amplifier frequency responses there are also deviations with respect to the signal display If relatively low frequency signals 20Hz are present the measurement value continuously change...

Страница 23: ...maximum symbol positioning and measuring accuracy first set the signal slope to the screen center X POS 12 and then activate X magnifier X MAG x10 13 To avoid CURSOR line and symbol changes after each...

Страница 24: ...reference value The measured value is calculated from the distance between the short CURSOR lines compared to the reference value This method is suitable to determine e g the oscilloscope s frequency...

Страница 25: ...as Y1 channel I Y2 channel II and M MATH signal their availability depends on the actual operating mode 1 If DUAL or XY mode is present in combination with CURSOR voltage measurement CURSOR MEASURE V...

Страница 26: ...m highlighted the equation to be displayed later can be chosen MAT1 MAT2 MAT3 and the mathematic function can be switched ON x or OFF by pressing the x pushbutton CAL SEL Pressing and holding opens th...

Страница 27: ...ith double function STOR ON Pressing and holding switches over from analog to storage digital mode and vice versa In the case of CT Component Tester mode active this mode must be left before it is pos...

Страница 28: ...right by the readout The following description assumes that the trigger conditions are met in Refresh rfr and it s submodes Envelope env and Average avm 41 1 1 rfr Refresh mode In this mode as in anal...

Страница 29: ...ease note the HO79 6 manual E MAIN MENU The instrument software contains several menus The following menus submenus and menu items are available 1 ADJUSTMENT contains 1 1 AUTO ADJUSTMENT with the item...

Страница 30: ...ode is not selected XY not displayed in the readout To obtain the maximum life from the cathode ray tube the minimum intensity setting necessary for the measurement in hand and the ambient light condi...

Страница 31: ...n should have the same value as during the 1kHz adjustment Probes other than those mentioned above normally have a larger tip diameter and may not fit into the calibrator output Whilst it is not diffi...

Страница 32: ...ue to phase shift do not use too high a test frequency It cannot be seen as a matter of course from the screen display if the test voltage leads or lags the reference voltage A CR network before the t...

Страница 33: ...r external trigger voltage A base line will always be displayed even with no signal With an applied AC signal peak value triggering enables the user to select the trigger point on the displayed signal...

Страница 34: ...er it is not signified by the trigger indicator LED which may still be blinking Break down of triggering is best recognizable at the left screen edge the start of the trace in differing display height...

Страница 35: ...frequency of the Y signal and is recommended for all mains line synchronous signals This also applies within certain limits to whole number multiples or fractions of the line frequency Line triggering...

Страница 36: ...is possible with certain pulse signals where the pulses alternately show a small difference of the peak amplitudes Only a very exact trigger level adjustment makes a single display possible The use o...

Страница 37: ...so possible in the dual channel addition and difference modes In chopped DUAL mode if after switching over to del or dTr the time deflection coefficient is reduced TIME DIV the channel switching mode...

Страница 38: ...ntrols and Readout The instrument has a built in electronic Component Tester which is used for instant display of a test pattern to indicate whether or not components are faulty It can be used for qui...

Страница 39: ...relatively small capacitance or a relatively high inductance A vertical ellipse indicates a low impedance or a relatively large capacitance or a relatively small inductance A sloping ellipse means tha...

Страница 40: ...e displayed as a complete curve Narrow pulses with low repetition rates do not cause intensity reduction Documentation and processing of captured signals is easily possible In comparison with analog m...

Страница 41: ...ired graticule position The trigger point symbol then should be set above or below the 0 Volt position line according to the expected voltage of the event to be captured Whether the slope selection is...

Страница 42: ...ting the sampling rate is too low the display of an alias signal may occur As described under Controls andReadout item 22 TIME DIV 22 3 AL thereadoutdisplays a warning if less than 2 samples are taken...

Страница 43: ...ching the oscilloscope off or transmitting the command RM 0 from the PC to the oscilloscope or depressing the AUTOSET LOCAL pushbutton if in unlocked condition command LK 1 was not sent After the remo...

Страница 44: ...uk Spain HAMEG S L Villarroel 172 174 08036 BARCELONA Tel f 93 4301597 Telefax 93 321220 E mail email hameg es France HAMEG S a r l 5 9 av de la R publique 94800 VILLEJUIF T l 1 4677 8151 Telefax 1 4...

Отзывы: