background image

24

Subject to change without notice

assigned to the short CURSOR line (SELECT (34)) which
must then be set to the pulse end position (4 div. after the
pulse start). Corresponding to the ratio of pulse duration
to period length (4:5 = 0.8) “0.8” will be displayed.

%:

Same function as described before under “Ratio”. The
measurement result is displayed in % (unit).

°

:

Angle measurement referring to the CURSOR line
distances. The distance between the long CURSOR lines
should cover one signal period, equivalent to 360°. Angle
measurement then can be performed by shifting the short
CURSOR line. Additional information can be found in
section “Operating modes of the vertical amplifiers in Yt
mode” under “Phase difference measurement in DUAL
mode”.

pi:

Determination of the value for “pi” referring to the
CURSOR line distances. The equivalent for “2 pi” is one
sine wave period; thus the distance between the long
CURSOR lines must be 1 period. If the distance between
the long CURSOR in left hand position and the short
CURSOR line referring to it is 1.5 periods, “3 pi” will be
indicated.

32.2.7 Ratio Y

 (display “ratio:Y, measured value, unit”)

Analog and digital mode.
Ratio Y measurement causes the display of two long and one
short CURSOR lines and is enabled only in Yt (time base) mode.

Briefly pressing the UNIT (35) pushbutton switches over
between the ratio (unnamed) and ratio in %.

The long CURSOR line in the lower position always serves
as the reference line. A “–“ (minus) sign indicates
measurement results if the short CURSOR line is placed
below  the reference line.

Ratio:

The distance between both long CURSOR lines is
equivalent to1.

Example: If the distance between the long CURSOR lines
is 6 div. and the short CURSOR line is activated (SELECT
(34)) and set 4 div. above the reference CURSOR line, the
ratio is 4:6, causing “0.667” (without unit) to be displayed

%:

The only difference between previous item “Ratio” and
“%” is that the distance between the long CURSOR lines
is equivalent to100% and the measuring result is displayed
as a % value.

32.2.8 Gain

 (display “gain: measured value, unit”)

Analog and digital mode.
Ratio measurement of signal voltages by the aid of two
long and two short CURSOR lines; enabled only in Yt (time
base) mode.

Briefly pressing the UNIT (35) pushbutton selects ratio
(unnamed), % or dB.

The application of Gain measurement depends whether
one or two signals are displayed.

1. One signal (CH I, CH II or “add” mode).
A measurement can be made on one signal before and
after a signal frequency change.

The distance between the long CURSOR lines serves as
the reference value. The measured value is calculated from
the distance between the short CURSOR lines compared
to the reference value.

This method is suitable to determine e.g. the oscilloscope‘s
frequency response.

2. DUAL mode.
Enables two port measurements (amplifier, attenuator) by
determination of the ratio of input and output voltages. For
correct measuring results you must determine which
channel is applied to the input and output ports respectively.

Both long CURSOR lines must be placed on the channel I
signal while the short CURSOR lines must be set on the
channel II signal.

Briefly pressing the SOURCE pushbutton calls a menu
which offers “g1

2:“ and “g2

1:“. The selection of the

required setting can then be made by briefly pressing the
SOURCE pushbutton until the setting is highlighted. If
channel I is connected to the input and channel II the output
of the two ports, “g1

2:“ must be chosen. Conversely if

the channels are reversed choose “g2

1:“.

32.2.9 rms

 (display “rms: channel, measured value”)

Digital mode only.
This function calculates and displays the rms (effective)
value of any signal period between the CURSOR lines. To
ensure that there is exactly one signal period between the
CURSOR lines, the SET (32) function can be called.

In cases where a signal consists of several signal periods
with different heights, the CURSOR lines must be set
manually to select the desired period. The measured value
is related to the signal part between the CURSOR lines,
which must be exactly one period.

DC contents are taken into account if DC input coupling is
present.

The “channel” to which the measurement result relates,
can be selected by the SOURCE (33) function. The
“channels” are indicated as “Y1” (channel I), “Y2” (channel
II) and “M” (MATH signal); their availability depends on the
actual operating mode.

32.2.10 avg

 (display “avg: channel, measured value”)

Digital mode only.
This measurement function calculates and displays the
average value of the signal within the CURSOR lines. DC
contents are taken into account if DC input coupling is
present.

The “channel” to which the measurement result relates,
can be selected by the SOURCE (33) function. The
“channels” are indicated as “Y1” (channel I), “Y2” (channel
II) and “M” (MATH signal); their availability depends on the
actual operating mode.

32.2.11 Peak Peak

 (display “pp: channel, measured value, unit”)

Digital mode only.
Automatically detects and marks (by triangle symbols) the
maximum amplitude difference in a sector determined by
two vertical CURSOR lines. The triangle symbols position
automatically and follow amplitude changes.

The UNIT (35) pushbutton can be used to switch over to
time difference measurement. Then the time difference
between the triangle symbols is displayed.

Controls and readout

Содержание HM507

Страница 1: ...Instruments Oscilloscope HM507 E N G L I S H HANDBUCH MANUAL MANUEL...

Страница 2: ...MANUAL HANDBUCH MANUEL...

Страница 3: ...tomatic Peak value Triggering 33 Normal Triggering 34 Slope 34 Trigger coupling 34 Triggering of video signals 34 Line Mains triggering 35 Alternate triggering 35 External triggering 35 Trigger indica...

Страница 4: ...place via the device under test mains line supply test leads control cables and or radiation The device under test as well as the oscilloscope may be effected by such fields Although the interior of t...

Страница 5: ...Interface 50MHz Analog Digital Oscilloscope HM507 Oscilloscope Specifications Vertical Deflection analog digital Operating modes Channel I or CH II separate CH I and II alternate or chopped 0 5MHz Sum...

Страница 6: ...be negated by the use of an extension cord without a protective conductor The mains line plug should be inserted before connections are made to measuring circuits The grounded accessible metal parts c...

Страница 7: ...data are based Purchase of the new HAMEG scope tester HZ 60 which despite its low price is highly suitable for tasks of this type is very much recommended The exterior of the oscilloscope should be cl...

Страница 8: ...lue is 11 higher The gain reduction error can not be defined exactly as the 3dB bandwidth of the amplifiers differ between 40MHz and 42MHz For sinewave signals the 6dB limit is approx 50MHz When exami...

Страница 9: ...se four values are not freely selectable They have to be within the following limits L between 0 2 and 10div if possible 4 to 10div T between 10ns and 5s F between 0 5Hz and 100MHz Tc between 100ns di...

Страница 10: ...ng the probe should always be used The series impedance of the probe provides a certain amount of protection for the input of the vertical amplifier Because of their separate manufacture all attenuato...

Страница 11: ...at 1 Component Tester is switched off 2 The following settings are present under MAIN MENU SETUP INFO MISCELLANEOUS 2 1 CONTROL BEEP and ERROR BEEP activated x 2 2 QUICK START not activated 3 The scre...

Страница 12: ...memories These can be used by the operator to save instrument settings and to recall them SAVE Press and hold the RECALL SAVE button to start a storage process This causes the SAVE menu Standard menu...

Страница 13: ...e determined by taking the deflection coefficient the probe factor and the trace position change with respect to the previous 0 Volt position into account 0 Volt Symbol The READOUT indicates the 0 Vol...

Страница 14: ...tomatic or automatic peak value triggering is selected AT Automatic triggering can be carried out with or without peak capture In both cases the LEVEL control 11 is effective and the trace is visible...

Страница 15: ...direction ccw The available range is from 1mV div up to 20V div The deflection coefficients and additional information regarding the active channel s are displayed in the readout e g Y1 deflection coe...

Страница 16: ...coefficient display as the time base is switched off 2 Controls and readout display switched off for trigger source slope and coupling 3 Y POS CURS I knob inactive for X shift use X POS 12 knob 4 X MA...

Страница 17: ...nel II and the READOUT displays Y2 trigger slope trigger coupling The last function setting of the VOLTS DIV 18 knob remains unchanged All channel II related controls are active if the input 28 is not...

Страница 18: ...zero and maximum can be set The signal displayed to the left of the trigger point symbol shows the pre history the right side displays the run of the curve after the trigger event POST TRIGGER As defi...

Страница 19: ...cient and can be selected by the TIME DIV knob It is recommended to capture and display the signals in DUAL mode and choose a suitable time coefficient which enables you to see the higher frequency si...

Страница 20: ...7 divisions The blanked section serves as a guide to the delay time The delay time is based on the current time deflection coefficient setting and can also be coarsely set with the TIME DIV control ra...

Страница 21: ...in conjunction with an operating mode instead of a measuring value the READOUT indicates n a not applicable For example the READOUT displays t n a if t measurement is selected in combination with XY m...

Страница 22: ...er and trigger amplifier frequency responses there are also deviations with respect to the signal display If relatively low frequency signals 20Hz are present the measurement value continuously change...

Страница 23: ...maximum symbol positioning and measuring accuracy first set the signal slope to the screen center X POS 12 and then activate X magnifier X MAG x10 13 To avoid CURSOR line and symbol changes after each...

Страница 24: ...reference value The measured value is calculated from the distance between the short CURSOR lines compared to the reference value This method is suitable to determine e g the oscilloscope s frequency...

Страница 25: ...as Y1 channel I Y2 channel II and M MATH signal their availability depends on the actual operating mode 1 If DUAL or XY mode is present in combination with CURSOR voltage measurement CURSOR MEASURE V...

Страница 26: ...m highlighted the equation to be displayed later can be chosen MAT1 MAT2 MAT3 and the mathematic function can be switched ON x or OFF by pressing the x pushbutton CAL SEL Pressing and holding opens th...

Страница 27: ...ith double function STOR ON Pressing and holding switches over from analog to storage digital mode and vice versa In the case of CT Component Tester mode active this mode must be left before it is pos...

Страница 28: ...right by the readout The following description assumes that the trigger conditions are met in Refresh rfr and it s submodes Envelope env and Average avm 41 1 1 rfr Refresh mode In this mode as in anal...

Страница 29: ...ease note the HO79 6 manual E MAIN MENU The instrument software contains several menus The following menus submenus and menu items are available 1 ADJUSTMENT contains 1 1 AUTO ADJUSTMENT with the item...

Страница 30: ...ode is not selected XY not displayed in the readout To obtain the maximum life from the cathode ray tube the minimum intensity setting necessary for the measurement in hand and the ambient light condi...

Страница 31: ...n should have the same value as during the 1kHz adjustment Probes other than those mentioned above normally have a larger tip diameter and may not fit into the calibrator output Whilst it is not diffi...

Страница 32: ...ue to phase shift do not use too high a test frequency It cannot be seen as a matter of course from the screen display if the test voltage leads or lags the reference voltage A CR network before the t...

Страница 33: ...r external trigger voltage A base line will always be displayed even with no signal With an applied AC signal peak value triggering enables the user to select the trigger point on the displayed signal...

Страница 34: ...er it is not signified by the trigger indicator LED which may still be blinking Break down of triggering is best recognizable at the left screen edge the start of the trace in differing display height...

Страница 35: ...frequency of the Y signal and is recommended for all mains line synchronous signals This also applies within certain limits to whole number multiples or fractions of the line frequency Line triggering...

Страница 36: ...is possible with certain pulse signals where the pulses alternately show a small difference of the peak amplitudes Only a very exact trigger level adjustment makes a single display possible The use o...

Страница 37: ...so possible in the dual channel addition and difference modes In chopped DUAL mode if after switching over to del or dTr the time deflection coefficient is reduced TIME DIV the channel switching mode...

Страница 38: ...ntrols and Readout The instrument has a built in electronic Component Tester which is used for instant display of a test pattern to indicate whether or not components are faulty It can be used for qui...

Страница 39: ...relatively small capacitance or a relatively high inductance A vertical ellipse indicates a low impedance or a relatively large capacitance or a relatively small inductance A sloping ellipse means tha...

Страница 40: ...e displayed as a complete curve Narrow pulses with low repetition rates do not cause intensity reduction Documentation and processing of captured signals is easily possible In comparison with analog m...

Страница 41: ...ired graticule position The trigger point symbol then should be set above or below the 0 Volt position line according to the expected voltage of the event to be captured Whether the slope selection is...

Страница 42: ...ting the sampling rate is too low the display of an alias signal may occur As described under Controls andReadout item 22 TIME DIV 22 3 AL thereadoutdisplays a warning if less than 2 samples are taken...

Страница 43: ...ching the oscilloscope off or transmitting the command RM 0 from the PC to the oscilloscope or depressing the AUTOSET LOCAL pushbutton if in unlocked condition command LK 1 was not sent After the remo...

Страница 44: ...uk Spain HAMEG S L Villarroel 172 174 08036 BARCELONA Tel f 93 4301597 Telefax 93 321220 E mail email hameg es France HAMEG S a r l 5 9 av de la R publique 94800 VILLEJUIF T l 1 4677 8151 Telefax 1 4...

Отзывы: