background image

8

Subject to change without notice

Voltage values of a sine curve

Vrms = effective value; Vp = simple peak or crest value;
Vpp = peak-to-peak value; Vmom = momentary value.

The minimum signal voltage which must be applied to the Y input
for a trace of 1div height is 1mVpp (± 5%) when this deflection
coefficient is displayed on the screen (readout) and the vernier is
switched off (VAR-LED dark). However, smaller signals than this
may also be displayed. The deflection coefficients are indicated
in mV/div or V/div (peak-to-peak value).

The magnitude of the applied voltage is ascertained by multiplying
the selected deflection coefficient by the vertical display height in
div. If an attenuator probe x10 is used, a further multiplication by
a factor of 10 is required to ascertain the correct voltage value.

For exact amplitude measurements, the variable control (VAR)
must be set to its calibrated detent CAL position.

With the variable control activated the deflection sensitivity
can be reduced up to a ratio of 2.5 to 1 (please note “controls
and readout”). Therefore any intermediate value is possible
within the 1-2-5 sequence of the attenuator(s).

With direct connection to the vertical input, sig-
nals up to 400Vpp may be displayed (attenuator
set to 20V/div, variable control to 2.5:1).

With the designations

H = display height in div,
U = signal  voltage in Vpp at the vertical input,
D = deflection coefficient in V/div at attenuator switch,

the required value can be calculated from the two given
quantities:

However, these three values are not freely selectable. They
have to be within the following limits (trigger threshold,
accuracy of reading):

H between 0.5 and 8div, if possible 3.2 to 8div,
U between 0.5mVpp and 160Vpp,
D between 1mV/div and 20V/div in 1-2-5 sequence.

Examples:
Set deflection coefficient D = 50mV/div   0.05V/div,
observed display height H = 4.6div,
required voltage U = 0.05x4.6 = 0.23Vpp.
Input voltage U = 5Vpp,
set deflection coefficient D = 1V/div,
required display height H = 5:1 = 5div.
Signal voltage U = 230Vrmsx2

√√√√√

2 = 651Vpp

(voltage 

160Vpp, with probe 10:1: U = 65.1Vpp),

desired display height H = min. 3.2div, max. 8div,

Type of signal voltage

The oscilloscope

 HM507 

allows examination of DC voltages

and most repetitive signals in the frequency range up to at least
40MHz (-3dB).

The vertical amplifiers have been designed for minimum
overshoot and therefore permit a true signal display.

The display of sinusoidal signals within the bandwidth limits
causes no problems, but an increasing error in measurement
due to gain reduction must be taken into account when
measuring high frequency signals. This error becomes
noticeable at approx. 14MHz. At approx. 18MHz the reduction
is approx. 10% and the real voltage value is 11% higher. The
gain reduction error can not be defined exactly as the -3dB
bandwidth of the amplifiers differ between 40MHz and 42MHz.
For sinewave signals the -6dB limit is approx. 50MHz.

When examining square or pulse type waveforms, attention
must be paid to the harmonic content of such signals. The
repetition frequency (fundamental frequency) of the signal
must therefore be significantly smaller than the upper limit
frequency of the vertical amplifier.

Displaying composite signals can be difficult, especially if they
contain no repetitive higher amplitude content which can be
used for triggering. This is the case with bursts, for instance.
To obtain a well-triggered display in this case, the assistance
of the variable holdoff function or the delayed time base may
be required. Television video signals are relatively easy to
trigger using the built-in TV-Sync-Separator (TV).

For optional operation as a DC or AC voltage amplifier, each
vertical amplifier input is provided with a DC/AC switch. DC
coupling should only be used with a series-connected attenuator
probe or at very low frequencies or if the measurement of the
DC voltage content of the signal is absolutely necessary.

When displaying very low frequency pulses, the flat tops may be
sloping with AC coupling of the vertical amplifier (AC limit frequency
approx. 1.6 Hz for 3dB). In this case, DC operation is preferred, provided
the signal voltage is not superimposed on a too high DC level.
Otherwise a capacitor of adequate capacitance must be connected
to the input of the vertical amplifier with DC coupling. This capacitor
must have a sufficiently high breakdown voltage rating. DC coupling
is also recommended for the display of logic and pulse signals,
especially if the pulse duty factor changes constantly. Otherwise the
display will move upwards or downwards at each change. Pure direct
voltages can only be measured with DC-coupling.

The input coupling is selectable by the AC/DC pushbutton. The
actual setting is displayed in the readout with the

 ” = ” 

symbol

for DC- and the

 ” 

~

 ” 

symbol for AC coupling.

Amplitude Measurements

In general electrical engineering, alternating voltage data normally
refers to effective values (rms = root-mean-square value). However,
for signal magnitudes and voltage designations in oscilloscope
measurements, the peak-to-peak voltage (Vpp) value is applied.
The latter corresponds to the real potential difference between the
most positive and most negative points of a signal waveform.

If a sinusoidal waveform, displayed on the oscilloscope screen,
is to be converted into an effective (rms) value, the resulting peak-
to-peak value must be divided by 2x

2 = 2.83. Conversely, it

should be observed that sinusoidal voltages indicated in Vrms
(Veff) have 2.83 times the potential difference in Vpp. The
relationship between the different voltage magnitudes can be
seen from the following figure.

Type of signal voltage

Содержание HM507

Страница 1: ...Instruments Oscilloscope HM507 E N G L I S H HANDBUCH MANUAL MANUEL...

Страница 2: ...MANUAL HANDBUCH MANUEL...

Страница 3: ...tomatic Peak value Triggering 33 Normal Triggering 34 Slope 34 Trigger coupling 34 Triggering of video signals 34 Line Mains triggering 35 Alternate triggering 35 External triggering 35 Trigger indica...

Страница 4: ...place via the device under test mains line supply test leads control cables and or radiation The device under test as well as the oscilloscope may be effected by such fields Although the interior of t...

Страница 5: ...Interface 50MHz Analog Digital Oscilloscope HM507 Oscilloscope Specifications Vertical Deflection analog digital Operating modes Channel I or CH II separate CH I and II alternate or chopped 0 5MHz Sum...

Страница 6: ...be negated by the use of an extension cord without a protective conductor The mains line plug should be inserted before connections are made to measuring circuits The grounded accessible metal parts c...

Страница 7: ...data are based Purchase of the new HAMEG scope tester HZ 60 which despite its low price is highly suitable for tasks of this type is very much recommended The exterior of the oscilloscope should be cl...

Страница 8: ...lue is 11 higher The gain reduction error can not be defined exactly as the 3dB bandwidth of the amplifiers differ between 40MHz and 42MHz For sinewave signals the 6dB limit is approx 50MHz When exami...

Страница 9: ...se four values are not freely selectable They have to be within the following limits L between 0 2 and 10div if possible 4 to 10div T between 10ns and 5s F between 0 5Hz and 100MHz Tc between 100ns di...

Страница 10: ...ng the probe should always be used The series impedance of the probe provides a certain amount of protection for the input of the vertical amplifier Because of their separate manufacture all attenuato...

Страница 11: ...at 1 Component Tester is switched off 2 The following settings are present under MAIN MENU SETUP INFO MISCELLANEOUS 2 1 CONTROL BEEP and ERROR BEEP activated x 2 2 QUICK START not activated 3 The scre...

Страница 12: ...memories These can be used by the operator to save instrument settings and to recall them SAVE Press and hold the RECALL SAVE button to start a storage process This causes the SAVE menu Standard menu...

Страница 13: ...e determined by taking the deflection coefficient the probe factor and the trace position change with respect to the previous 0 Volt position into account 0 Volt Symbol The READOUT indicates the 0 Vol...

Страница 14: ...tomatic or automatic peak value triggering is selected AT Automatic triggering can be carried out with or without peak capture In both cases the LEVEL control 11 is effective and the trace is visible...

Страница 15: ...direction ccw The available range is from 1mV div up to 20V div The deflection coefficients and additional information regarding the active channel s are displayed in the readout e g Y1 deflection coe...

Страница 16: ...coefficient display as the time base is switched off 2 Controls and readout display switched off for trigger source slope and coupling 3 Y POS CURS I knob inactive for X shift use X POS 12 knob 4 X MA...

Страница 17: ...nel II and the READOUT displays Y2 trigger slope trigger coupling The last function setting of the VOLTS DIV 18 knob remains unchanged All channel II related controls are active if the input 28 is not...

Страница 18: ...zero and maximum can be set The signal displayed to the left of the trigger point symbol shows the pre history the right side displays the run of the curve after the trigger event POST TRIGGER As defi...

Страница 19: ...cient and can be selected by the TIME DIV knob It is recommended to capture and display the signals in DUAL mode and choose a suitable time coefficient which enables you to see the higher frequency si...

Страница 20: ...7 divisions The blanked section serves as a guide to the delay time The delay time is based on the current time deflection coefficient setting and can also be coarsely set with the TIME DIV control ra...

Страница 21: ...in conjunction with an operating mode instead of a measuring value the READOUT indicates n a not applicable For example the READOUT displays t n a if t measurement is selected in combination with XY m...

Страница 22: ...er and trigger amplifier frequency responses there are also deviations with respect to the signal display If relatively low frequency signals 20Hz are present the measurement value continuously change...

Страница 23: ...maximum symbol positioning and measuring accuracy first set the signal slope to the screen center X POS 12 and then activate X magnifier X MAG x10 13 To avoid CURSOR line and symbol changes after each...

Страница 24: ...reference value The measured value is calculated from the distance between the short CURSOR lines compared to the reference value This method is suitable to determine e g the oscilloscope s frequency...

Страница 25: ...as Y1 channel I Y2 channel II and M MATH signal their availability depends on the actual operating mode 1 If DUAL or XY mode is present in combination with CURSOR voltage measurement CURSOR MEASURE V...

Страница 26: ...m highlighted the equation to be displayed later can be chosen MAT1 MAT2 MAT3 and the mathematic function can be switched ON x or OFF by pressing the x pushbutton CAL SEL Pressing and holding opens th...

Страница 27: ...ith double function STOR ON Pressing and holding switches over from analog to storage digital mode and vice versa In the case of CT Component Tester mode active this mode must be left before it is pos...

Страница 28: ...right by the readout The following description assumes that the trigger conditions are met in Refresh rfr and it s submodes Envelope env and Average avm 41 1 1 rfr Refresh mode In this mode as in anal...

Страница 29: ...ease note the HO79 6 manual E MAIN MENU The instrument software contains several menus The following menus submenus and menu items are available 1 ADJUSTMENT contains 1 1 AUTO ADJUSTMENT with the item...

Страница 30: ...ode is not selected XY not displayed in the readout To obtain the maximum life from the cathode ray tube the minimum intensity setting necessary for the measurement in hand and the ambient light condi...

Страница 31: ...n should have the same value as during the 1kHz adjustment Probes other than those mentioned above normally have a larger tip diameter and may not fit into the calibrator output Whilst it is not diffi...

Страница 32: ...ue to phase shift do not use too high a test frequency It cannot be seen as a matter of course from the screen display if the test voltage leads or lags the reference voltage A CR network before the t...

Страница 33: ...r external trigger voltage A base line will always be displayed even with no signal With an applied AC signal peak value triggering enables the user to select the trigger point on the displayed signal...

Страница 34: ...er it is not signified by the trigger indicator LED which may still be blinking Break down of triggering is best recognizable at the left screen edge the start of the trace in differing display height...

Страница 35: ...frequency of the Y signal and is recommended for all mains line synchronous signals This also applies within certain limits to whole number multiples or fractions of the line frequency Line triggering...

Страница 36: ...is possible with certain pulse signals where the pulses alternately show a small difference of the peak amplitudes Only a very exact trigger level adjustment makes a single display possible The use o...

Страница 37: ...so possible in the dual channel addition and difference modes In chopped DUAL mode if after switching over to del or dTr the time deflection coefficient is reduced TIME DIV the channel switching mode...

Страница 38: ...ntrols and Readout The instrument has a built in electronic Component Tester which is used for instant display of a test pattern to indicate whether or not components are faulty It can be used for qui...

Страница 39: ...relatively small capacitance or a relatively high inductance A vertical ellipse indicates a low impedance or a relatively large capacitance or a relatively small inductance A sloping ellipse means tha...

Страница 40: ...e displayed as a complete curve Narrow pulses with low repetition rates do not cause intensity reduction Documentation and processing of captured signals is easily possible In comparison with analog m...

Страница 41: ...ired graticule position The trigger point symbol then should be set above or below the 0 Volt position line according to the expected voltage of the event to be captured Whether the slope selection is...

Страница 42: ...ting the sampling rate is too low the display of an alias signal may occur As described under Controls andReadout item 22 TIME DIV 22 3 AL thereadoutdisplays a warning if less than 2 samples are taken...

Страница 43: ...ching the oscilloscope off or transmitting the command RM 0 from the PC to the oscilloscope or depressing the AUTOSET LOCAL pushbutton if in unlocked condition command LK 1 was not sent After the remo...

Страница 44: ...uk Spain HAMEG S L Villarroel 172 174 08036 BARCELONA Tel f 93 4301597 Telefax 93 321220 E mail email hameg es France HAMEG S a r l 5 9 av de la R publique 94800 VILLEJUIF T l 1 4677 8151 Telefax 1 4...

Отзывы: