background image

35

Subject to change without notice

Video signals are triggered in the automatic mode. The internal
triggering is virtually independent of the display height, but
the sync pulse must exceed 0.5div height.

The polarity of the synchronization pulse is critical for the slope
selection. If the displayed sync pulses are above the picture
(field) contents (leading edge positive going), then the slope
setting for positive going edges must be chosen. In the case
of sync pulses below the field/line, the leading edge is negative
and consequently the slope selection must be set for falling
edges. Since the invert function may cause a misleading
display, it must not be activated.

On the 2ms/div setting and field     

TV

     triggering selected, 1

field is visible if a 50 fields/s signal is applied. If the hold off
control is in fully ccw position, it triggers without line interlacing
affects caused by the consecutive field.

The display can be expanded by switching on the     

X-MAG.

x10

 function so that individual lines are recognizable.

Commencing with a frame synchronizing pulse, the display
can also be expanded with the knob     

TIME/DIV

. But note that

this  can result in an apparently unsynchronized display as
each frame (half picture) triggers. This is due to the off set of
half a line between frames.

The influence of the integrating network which forms a trigger
pulse from the vertical sync pulses may become visible under
certain conditions. Due to the integrating network time constant
not all vertical sync pulses starting the trace are visible.

On the 10µs/div setting and line TV triggering selected, approx.
1½  lines are visible. Those lines originate from the odd and
even fields at random.

The sync-separator-circuit also operates with external
triggering. It is important that the voltage range (0.3Vpp to
3Vpp) for external triggering should be noted. Again the correct
slope setting is critical, because the external trigger signal
may not have the same polarity or pulse edge as the test signal
displayed on the CRT. This can be checked, if the external
trigger voltage itself is displayed first (with internal triggering).

In most cases, the composite video signal has a high DC
content. With constant video information (e.g. test pattern or
colour bar generator), the DC content can be suppressed easily
by AC input coupling of the oscilloscope amplifier. With a
changing picture content (e.g. normal program), DC input
coupling is recommen-ded, because the display varies its
vertical position on screen with AC input coupling at each
change of the picture content. The DC content can be
compensated using the     

Y POS

. control so that the signal display

lies in the graticule area. Then the composite video signal
should not exceed a vertical height of 6div.

Line/Mains triggering (

~

)

The instrument specific information regarding this
mode is part of the section ”Controls and Readout”
paragraph TRIG. MODE (20).

This trigger mode is present if the READOUT indicates the
“~” 

“~” 

“~” 

“~” 

“~” symbol instead of the “trigger source”,  “slope” and
“coupling” information. The trigger point symbol is inactive in
line/mains trigger mode as there is no direct amplitude
relationship between the trigger voltage and the signal voltage.

A voltage originating from mains/line (50 to 60Hz) is used for
triggering purposes if the trigger coupling is set to ~

~

~

~

~. This

trigger mode is independent of amplitude and frequency of
the Y signal and is recommended for all mains/line synchronous

signals. This also applies  within certain limits,  to whole number
multiples or fractions of the line frequency. Line triggering can
also be useful to display signals below the trigger threshold
(less than 0.5div). It is therefore particularly suitable for
measuring small ripple voltages of mains/line rectifiers or stray
magnetic field in a circuit. In this trigger mode the slope
direction pushbutton selects the positive or negative portion
of the line/mains

 

sinewave. The trigger level control can be

used for trigger point adjustment.

Magnetic leakage (e.g. from a power transformer) can be
investigated for direction and amplitude using a search or pick
up coil. The coil should be wound on a small former with a
maximum of turns of a thin lacquered wire and connected to a
BNC connector (for scope input) via a shielded cable. Between
cable and BNC center conductor a resistor of at least 100 Ohm
should be series connected (RF decoupling). Often it is
advisable to statically shield the surface of the coil. However,
no shorted turns are permissible. Maximum, minimum, and
direction to the magnetic source are detectable at the
measuring point by turning and shifting the coil.

Alternate triggering

This trigger mode can be selected in DUAL mode by the aid of
the TRIG. SOURCE (17) pushbutton (please note ”Controls
and Readout”). In the case of chopped DUAL mode, selecting
alternate trigger mode automatically sets the instrument to
alternate DUAL mode.

Under TV-L, TV-F and line/mains triggering conditions alternate
triggering can not be chosen. Thus only the following trigger
coupling modes are available in alternate trigger mode: AC,
DC, HF and LF. The trigger point symbol is not displayed in
alternate trigger mode.

With alternate triggering it is possible to trigger two signals
from different frequency sources (asynchronous). In this case
the oscilloscope must be operated in DUAL     alternate mode
with internal triggering and each input signal must be of
sufficient height to enable trigger. To avoid trigger problems
due to different     DC     voltage components,     AC     input coupling
for both channels is recommended.

The internal trigger source is switched in alternate trigger mode
in the same way as the channel switching system in DUAL
alternate mode, i.e. after each time base sweep. Phase
difference measurement is not possible in this trigger mode
as the trigger level and slope setting are equal for both signals.
Even with 180° phase difference between both signals, they
appear with the same slope direction.

If signals are applied with a high frequency ratio (difference),
the trace intensity then becomes reduced if the time base is
set to smaller time coefficients (faster sweep). This happens
as the number of sweeps does not increase because it
depends on the lower frequency signal, but with a faster sweep
the phosphor becomes less activated.

External triggering

The external trigger input is activated with the aid of the     TRIG.
SOURCE (17) pushbutton (see ”Controls and Readout”), if the
trigger coupling is not set to line/mains trigger coupling. Then
the internal trigger source is deactivated. As the external trigger
signal applied at the T

 T

 T

 T

 TRIG. EXT socket normally has no relation

to the signal height of the displayed signal, the trigger point
symbol is switched off. The external trigger voltage must have
a minimum amplitude of 0.3Vpp and should not increase above
3Vpp. The input impedance of the     TRIG. EXT. socket is approx.
1M Ohm II 20pF.

Triggering and time base

Содержание HM507

Страница 1: ...Instruments Oscilloscope HM507 E N G L I S H HANDBUCH MANUAL MANUEL...

Страница 2: ...MANUAL HANDBUCH MANUEL...

Страница 3: ...tomatic Peak value Triggering 33 Normal Triggering 34 Slope 34 Trigger coupling 34 Triggering of video signals 34 Line Mains triggering 35 Alternate triggering 35 External triggering 35 Trigger indica...

Страница 4: ...place via the device under test mains line supply test leads control cables and or radiation The device under test as well as the oscilloscope may be effected by such fields Although the interior of t...

Страница 5: ...Interface 50MHz Analog Digital Oscilloscope HM507 Oscilloscope Specifications Vertical Deflection analog digital Operating modes Channel I or CH II separate CH I and II alternate or chopped 0 5MHz Sum...

Страница 6: ...be negated by the use of an extension cord without a protective conductor The mains line plug should be inserted before connections are made to measuring circuits The grounded accessible metal parts c...

Страница 7: ...data are based Purchase of the new HAMEG scope tester HZ 60 which despite its low price is highly suitable for tasks of this type is very much recommended The exterior of the oscilloscope should be cl...

Страница 8: ...lue is 11 higher The gain reduction error can not be defined exactly as the 3dB bandwidth of the amplifiers differ between 40MHz and 42MHz For sinewave signals the 6dB limit is approx 50MHz When exami...

Страница 9: ...se four values are not freely selectable They have to be within the following limits L between 0 2 and 10div if possible 4 to 10div T between 10ns and 5s F between 0 5Hz and 100MHz Tc between 100ns di...

Страница 10: ...ng the probe should always be used The series impedance of the probe provides a certain amount of protection for the input of the vertical amplifier Because of their separate manufacture all attenuato...

Страница 11: ...at 1 Component Tester is switched off 2 The following settings are present under MAIN MENU SETUP INFO MISCELLANEOUS 2 1 CONTROL BEEP and ERROR BEEP activated x 2 2 QUICK START not activated 3 The scre...

Страница 12: ...memories These can be used by the operator to save instrument settings and to recall them SAVE Press and hold the RECALL SAVE button to start a storage process This causes the SAVE menu Standard menu...

Страница 13: ...e determined by taking the deflection coefficient the probe factor and the trace position change with respect to the previous 0 Volt position into account 0 Volt Symbol The READOUT indicates the 0 Vol...

Страница 14: ...tomatic or automatic peak value triggering is selected AT Automatic triggering can be carried out with or without peak capture In both cases the LEVEL control 11 is effective and the trace is visible...

Страница 15: ...direction ccw The available range is from 1mV div up to 20V div The deflection coefficients and additional information regarding the active channel s are displayed in the readout e g Y1 deflection coe...

Страница 16: ...coefficient display as the time base is switched off 2 Controls and readout display switched off for trigger source slope and coupling 3 Y POS CURS I knob inactive for X shift use X POS 12 knob 4 X MA...

Страница 17: ...nel II and the READOUT displays Y2 trigger slope trigger coupling The last function setting of the VOLTS DIV 18 knob remains unchanged All channel II related controls are active if the input 28 is not...

Страница 18: ...zero and maximum can be set The signal displayed to the left of the trigger point symbol shows the pre history the right side displays the run of the curve after the trigger event POST TRIGGER As defi...

Страница 19: ...cient and can be selected by the TIME DIV knob It is recommended to capture and display the signals in DUAL mode and choose a suitable time coefficient which enables you to see the higher frequency si...

Страница 20: ...7 divisions The blanked section serves as a guide to the delay time The delay time is based on the current time deflection coefficient setting and can also be coarsely set with the TIME DIV control ra...

Страница 21: ...in conjunction with an operating mode instead of a measuring value the READOUT indicates n a not applicable For example the READOUT displays t n a if t measurement is selected in combination with XY m...

Страница 22: ...er and trigger amplifier frequency responses there are also deviations with respect to the signal display If relatively low frequency signals 20Hz are present the measurement value continuously change...

Страница 23: ...maximum symbol positioning and measuring accuracy first set the signal slope to the screen center X POS 12 and then activate X magnifier X MAG x10 13 To avoid CURSOR line and symbol changes after each...

Страница 24: ...reference value The measured value is calculated from the distance between the short CURSOR lines compared to the reference value This method is suitable to determine e g the oscilloscope s frequency...

Страница 25: ...as Y1 channel I Y2 channel II and M MATH signal their availability depends on the actual operating mode 1 If DUAL or XY mode is present in combination with CURSOR voltage measurement CURSOR MEASURE V...

Страница 26: ...m highlighted the equation to be displayed later can be chosen MAT1 MAT2 MAT3 and the mathematic function can be switched ON x or OFF by pressing the x pushbutton CAL SEL Pressing and holding opens th...

Страница 27: ...ith double function STOR ON Pressing and holding switches over from analog to storage digital mode and vice versa In the case of CT Component Tester mode active this mode must be left before it is pos...

Страница 28: ...right by the readout The following description assumes that the trigger conditions are met in Refresh rfr and it s submodes Envelope env and Average avm 41 1 1 rfr Refresh mode In this mode as in anal...

Страница 29: ...ease note the HO79 6 manual E MAIN MENU The instrument software contains several menus The following menus submenus and menu items are available 1 ADJUSTMENT contains 1 1 AUTO ADJUSTMENT with the item...

Страница 30: ...ode is not selected XY not displayed in the readout To obtain the maximum life from the cathode ray tube the minimum intensity setting necessary for the measurement in hand and the ambient light condi...

Страница 31: ...n should have the same value as during the 1kHz adjustment Probes other than those mentioned above normally have a larger tip diameter and may not fit into the calibrator output Whilst it is not diffi...

Страница 32: ...ue to phase shift do not use too high a test frequency It cannot be seen as a matter of course from the screen display if the test voltage leads or lags the reference voltage A CR network before the t...

Страница 33: ...r external trigger voltage A base line will always be displayed even with no signal With an applied AC signal peak value triggering enables the user to select the trigger point on the displayed signal...

Страница 34: ...er it is not signified by the trigger indicator LED which may still be blinking Break down of triggering is best recognizable at the left screen edge the start of the trace in differing display height...

Страница 35: ...frequency of the Y signal and is recommended for all mains line synchronous signals This also applies within certain limits to whole number multiples or fractions of the line frequency Line triggering...

Страница 36: ...is possible with certain pulse signals where the pulses alternately show a small difference of the peak amplitudes Only a very exact trigger level adjustment makes a single display possible The use o...

Страница 37: ...so possible in the dual channel addition and difference modes In chopped DUAL mode if after switching over to del or dTr the time deflection coefficient is reduced TIME DIV the channel switching mode...

Страница 38: ...ntrols and Readout The instrument has a built in electronic Component Tester which is used for instant display of a test pattern to indicate whether or not components are faulty It can be used for qui...

Страница 39: ...relatively small capacitance or a relatively high inductance A vertical ellipse indicates a low impedance or a relatively large capacitance or a relatively small inductance A sloping ellipse means tha...

Страница 40: ...e displayed as a complete curve Narrow pulses with low repetition rates do not cause intensity reduction Documentation and processing of captured signals is easily possible In comparison with analog m...

Страница 41: ...ired graticule position The trigger point symbol then should be set above or below the 0 Volt position line according to the expected voltage of the event to be captured Whether the slope selection is...

Страница 42: ...ting the sampling rate is too low the display of an alias signal may occur As described under Controls andReadout item 22 TIME DIV 22 3 AL thereadoutdisplays a warning if less than 2 samples are taken...

Страница 43: ...ching the oscilloscope off or transmitting the command RM 0 from the PC to the oscilloscope or depressing the AUTOSET LOCAL pushbutton if in unlocked condition command LK 1 was not sent After the remo...

Страница 44: ...uk Spain HAMEG S L Villarroel 172 174 08036 BARCELONA Tel f 93 4301597 Telefax 93 321220 E mail email hameg es France HAMEG S a r l 5 9 av de la R publique 94800 VILLEJUIF T l 1 4677 8151 Telefax 1 4...

Отзывы: