9-2
Microprocessor
P/N 108755-US, Manual revision: AA
B
ODY
B
OARD
R
EPLACEMENT
Warning: With the removal of the Body Board and Power Supply PCBA's from
the base, AC and DC voltages are exposed, use CAUTION when handling these
assemblies.
Caution:
Always wear an ESD strap when working on the Shaker.
1.
Remove body board and power supply from the base of the Shaker (see
Chapter 5, Body Board Strain Gauge Amp Adjustment Procedure on page
5-7).
2.
Disconnect all plug connectors from Body Board.
3.
Remove the four mounting screws and install the new Body Board.
4.
Attach the extender harness.
5.
Apply power to Shaker. Adjust R3 and R7 (see Chapter 5, Body Board
Strain Gauge Amp Adjustment Procedure on page 5-7).
6.
Verify all settings (see below).
7.
After installing a new body board, the unit must be programmed using the
instructions in Program setting "F1" on page 9-3 and Clock/Calendar Ini-
tialization on page 9-7.
Содержание SEBRA 1040
Страница 1: ...108755 US service manual Model 1040 Digital Blood Shaker...
Страница 3: ......
Страница 4: ...Printed in USA Copyright 1999 2010 Haemonetics Corporation...
Страница 8: ...vi Preface P N 108755 US Manual revision AA...
Страница 12: ...x Table of Contents P N 108755 US Manual revision AA...
Страница 14: ...xii List of Figures P N 108755 US Manual revision AA...
Страница 20: ...1 6 Introduction P N 108755 US Manual revision AA...
Страница 36: ...3 10 Inspection Procedures P N 108755 US Manual revision AA...
Страница 37: ...P N 108755 US Manual revision AA Chapter 4 Parts Access BASE PLATE REMOVAL 4 2 SHROUD REMOVAL 4 3...
Страница 40: ...4 4 Parts Access P N 108755 US Manual revision AA...
Страница 48: ...5 8 Strain Gauge P N 108755 US Manual revision AA...
Страница 60: ...7 6 Tube Clamp Assembly P N 108755 US Manual revision AA...
Страница 64: ...8 4 PCB Chassis Assembly Installation P N 108755 US Manual revision AA...
Страница 83: ...P N 108755 US Manual revision AA Chapter 11 Clamp Body Boss Replacement REPLACING THE BODY BOSS 11 2...
Страница 86: ...11 4 Clamp Body Boss Replacement P N 108755 US Manual revision AA...
Страница 98: ...13 8 Cap Assembly P N 108755 US Manual revision AA...
Страница 102: ...14 4 Power Supply P N 108755 US Manual revision AA...
Страница 103: ...P N 108755 US Manual revision AA Chapter 15 Shaker Keystroke Quick Reference KEYSTROKE QUICK REFERENCE 15 2...
Страница 105: ......
Страница 106: ...Haemonetics Corporation 400 Wood Road Braintree MA 02184 USA www haemonetics com 1995 2010 Haemonetics Corporation...