7-4
Tube Clamp Assembly
P/N 108755-US, Manual revision: AA
15.
Use light sand paper to remove build up on tube clamp. Assemble sole-
noid replacing the spacer. Remove residual super glue on shaft. Set the
length of the tube clamp and shaft to 4.125 + 0.010 - 0.005in.
16.
Push up on bottom of solenoid, hold spacer with thumb, tip solenoid up-
side down and apply super glue to the threads where the tube clamp
meets the solenoid shaft. Do not allow super glue to run the opposite di-
rection into the shaft of the solenoid. After glue seeps in between the
threads, drop the spacer to glue it to the tube clamp.
17.
Apply Dow Corning
®
55 grease Molycoat to the O-ring and tube clamp
with finger. Try to avoid greasing the face of the tube clamp.
18.
Reassemble tube clamp by placing the nose of the solenoid into tube
clamp and the tube clamp into tube holder with the wires facing the open-
ing. Fasten with the self-tapping screws without a washer. Do not tighten
completely until the other screw has been started. Adjust so it does not
bind and moves freely in 360
°
.
19.
Plug in the Shaker and place a blood bag set containing water on the bag
tray, with tubing threaded in the tube clamp. Press the <-> button to en-
ergize the solenoid and force an air bubble through the tubing. No move-
ment of air bubble through the tube clamp should be observed. Unplug
the Shaker.
Содержание SEBRA 1040
Страница 1: ...108755 US service manual Model 1040 Digital Blood Shaker...
Страница 3: ......
Страница 4: ...Printed in USA Copyright 1999 2010 Haemonetics Corporation...
Страница 8: ...vi Preface P N 108755 US Manual revision AA...
Страница 12: ...x Table of Contents P N 108755 US Manual revision AA...
Страница 14: ...xii List of Figures P N 108755 US Manual revision AA...
Страница 20: ...1 6 Introduction P N 108755 US Manual revision AA...
Страница 36: ...3 10 Inspection Procedures P N 108755 US Manual revision AA...
Страница 37: ...P N 108755 US Manual revision AA Chapter 4 Parts Access BASE PLATE REMOVAL 4 2 SHROUD REMOVAL 4 3...
Страница 40: ...4 4 Parts Access P N 108755 US Manual revision AA...
Страница 48: ...5 8 Strain Gauge P N 108755 US Manual revision AA...
Страница 60: ...7 6 Tube Clamp Assembly P N 108755 US Manual revision AA...
Страница 64: ...8 4 PCB Chassis Assembly Installation P N 108755 US Manual revision AA...
Страница 83: ...P N 108755 US Manual revision AA Chapter 11 Clamp Body Boss Replacement REPLACING THE BODY BOSS 11 2...
Страница 86: ...11 4 Clamp Body Boss Replacement P N 108755 US Manual revision AA...
Страница 98: ...13 8 Cap Assembly P N 108755 US Manual revision AA...
Страница 102: ...14 4 Power Supply P N 108755 US Manual revision AA...
Страница 103: ...P N 108755 US Manual revision AA Chapter 15 Shaker Keystroke Quick Reference KEYSTROKE QUICK REFERENCE 15 2...
Страница 105: ......
Страница 106: ...Haemonetics Corporation 400 Wood Road Braintree MA 02184 USA www haemonetics com 1995 2010 Haemonetics Corporation...