Inspection Procedures
3-5
P/N 108755-US, Manual revision: AA
C
ALIBRATION
Warning: Calibration must be performed according to the following approved
procedure using a Haemonetics
®
SEBRA
®
certified calibration weight. If the
Shaker is calibrated using any other weight or procedure, the accuracy of the
blood collection volume cannot be assured.
The Shaker should be calibrated after any repairs have been performed.
1.
From the "Ready" mode, press <
1
>. The display will read "pw".
2.
Enter the current password, then press <*>. The display will now read "set
up".
3.
Press <*>. The display will now read "Calibrat".
Note: If you accidentally get into the manual mode, the display will read "manu-
al". Immediately press <#> to return to "ready".
4.
Press <+>. The display will now read "zero wt".
Note: At this point, be sure there is nothing on, or resting against, the Shaker
shroud assembly.
5.
Press <*>. The shroud assembly will start oscillating. The display will first
read "weighing" for a short time, then the display will show a "
zerovalue
"
number. This number represents a minimum weight value allowed within
the calibration parameters. The shroud assembly will automatically stop
oscillating and the "
zerovalue
" number will be replaced by the words
"place wt".
Note: If for some reason it is necessary to start the calibration again at this point,
press <#> and the display will read "Calibrat". Go back to step 4 to try the cali-
bration again. If the display reads "FAILURE", be sure the Shaker is level and that
nothing is resting on or against the shroud assembly. Press any key on the keypad
to return to "Calibrat", and then go back to step 4 to try again.
6.
Place a Model 1024 (585 gram) certified calibration weight on the bag
tray, secured to the two bag pegs, with the hinged counterweight pointed
toward the pegs, as shown in Figure 3-1, Calibration Weight Placement.
Содержание SEBRA 1040
Страница 1: ...108755 US service manual Model 1040 Digital Blood Shaker...
Страница 3: ......
Страница 4: ...Printed in USA Copyright 1999 2010 Haemonetics Corporation...
Страница 8: ...vi Preface P N 108755 US Manual revision AA...
Страница 12: ...x Table of Contents P N 108755 US Manual revision AA...
Страница 14: ...xii List of Figures P N 108755 US Manual revision AA...
Страница 20: ...1 6 Introduction P N 108755 US Manual revision AA...
Страница 36: ...3 10 Inspection Procedures P N 108755 US Manual revision AA...
Страница 37: ...P N 108755 US Manual revision AA Chapter 4 Parts Access BASE PLATE REMOVAL 4 2 SHROUD REMOVAL 4 3...
Страница 40: ...4 4 Parts Access P N 108755 US Manual revision AA...
Страница 48: ...5 8 Strain Gauge P N 108755 US Manual revision AA...
Страница 60: ...7 6 Tube Clamp Assembly P N 108755 US Manual revision AA...
Страница 64: ...8 4 PCB Chassis Assembly Installation P N 108755 US Manual revision AA...
Страница 83: ...P N 108755 US Manual revision AA Chapter 11 Clamp Body Boss Replacement REPLACING THE BODY BOSS 11 2...
Страница 86: ...11 4 Clamp Body Boss Replacement P N 108755 US Manual revision AA...
Страница 98: ...13 8 Cap Assembly P N 108755 US Manual revision AA...
Страница 102: ...14 4 Power Supply P N 108755 US Manual revision AA...
Страница 103: ...P N 108755 US Manual revision AA Chapter 15 Shaker Keystroke Quick Reference KEYSTROKE QUICK REFERENCE 15 2...
Страница 105: ......
Страница 106: ...Haemonetics Corporation 400 Wood Road Braintree MA 02184 USA www haemonetics com 1995 2010 Haemonetics Corporation...