Strain Gauge
5-5
P/N 108755-US, Manual revision: AA
Figure 5-5, Oscilloscope Wave Form
11.
Stop Shaker weighing routine from keypad. Remove power from Shaker.
Detach adjusted assembly from fixture.
Figure 5-6, Weight Sensor Assembly-Current
12.
For those PCBAs at Revision G or higher, after step 5 attach a DVM ground
lead (black) to TP5 and the positive lead (red) to TP2 on the body board.
13.
Apply power to the Shaker. Press the "*" on the keypad to start the Shaker.
14.
With no weight on the knife edge of the strain gage assembly, the DVM
should read between 4.40 and 5.00 VDC.
15.
Hang the strain gage calibration weight (p/n 10401200-00) on the knife
edge, the DVM should read between 6.10 and 7.10 VDC.
16.
Remove the calibration weight, stop the Shaker and remove power. The
weight sensor verification is now complete.
Содержание SEBRA 1040
Страница 1: ...108755 US service manual Model 1040 Digital Blood Shaker...
Страница 3: ......
Страница 4: ...Printed in USA Copyright 1999 2010 Haemonetics Corporation...
Страница 8: ...vi Preface P N 108755 US Manual revision AA...
Страница 12: ...x Table of Contents P N 108755 US Manual revision AA...
Страница 14: ...xii List of Figures P N 108755 US Manual revision AA...
Страница 20: ...1 6 Introduction P N 108755 US Manual revision AA...
Страница 36: ...3 10 Inspection Procedures P N 108755 US Manual revision AA...
Страница 37: ...P N 108755 US Manual revision AA Chapter 4 Parts Access BASE PLATE REMOVAL 4 2 SHROUD REMOVAL 4 3...
Страница 40: ...4 4 Parts Access P N 108755 US Manual revision AA...
Страница 48: ...5 8 Strain Gauge P N 108755 US Manual revision AA...
Страница 60: ...7 6 Tube Clamp Assembly P N 108755 US Manual revision AA...
Страница 64: ...8 4 PCB Chassis Assembly Installation P N 108755 US Manual revision AA...
Страница 83: ...P N 108755 US Manual revision AA Chapter 11 Clamp Body Boss Replacement REPLACING THE BODY BOSS 11 2...
Страница 86: ...11 4 Clamp Body Boss Replacement P N 108755 US Manual revision AA...
Страница 98: ...13 8 Cap Assembly P N 108755 US Manual revision AA...
Страница 102: ...14 4 Power Supply P N 108755 US Manual revision AA...
Страница 103: ...P N 108755 US Manual revision AA Chapter 15 Shaker Keystroke Quick Reference KEYSTROKE QUICK REFERENCE 15 2...
Страница 105: ......
Страница 106: ...Haemonetics Corporation 400 Wood Road Braintree MA 02184 USA www haemonetics com 1995 2010 Haemonetics Corporation...