5. Display and Printing
5-1. Arterial and venous blood velocity modes
Arterial and venous blood velocity can be measured on these modes by Doppler
technique. Smartdop displays and prints waveform as follows:
Mode 1:
Displays approx. 2.56 sec of waveform per screen of LCD
Prints waveform in time scale of approx. 25 mm/sec.
Forward and reverse components are combined.
Mode 2:
Displays approx. 12.8 sec of waveforms per screen of LCD
Prints waveforms in time scale of approx. 5 mm/sec.
Forward component (arterial) is separated from reverse component (venous).
For typical operation for blood velocity modes, see “§ 3-5. Measuring Blood Velocity”.
5-1-1. Waveforms on LCD
The base line is automatically located at best position depending on how big the
reverse component is. Smartdop has 4 base lines, the bottom, 1/4 from the bottom, the
center, and 3/4 from the bottom.
The waveform amplitude is automatically adjusted for optimal observation.
The velocity amplitude scale per division is shown at top left of LCD.
Memory number is shown at top right of LCD for waveform read on MEMORY mode.
e.g. "#01"
Memory number
Amplitude
scale
Di
visi
o
n
Mode 1
Mode 2
Содержание Smartdop 30EX