12
(6) Headset can be used to listen to Doppler sounds. It
will cut off the speaker.
If you wish to save the waveform and numerical
data on the memory, see Ҥ 4-2-7-a. MEMORY -
STORE”.
If you retry the measurement, press INF/DUMP
button or probe button to get out of the freeze
mode.
3-6. Ankle brachial pressure index (ABI)
(1) Connect the cuff VC-12 to the tubing and other end
of tubing to the inflation inlet as shown in the right.
(2) Connect the probe to the probe connector #1 or
#2.
See the step# (1) in the “§.3-1. Turning the unit ON” for the details.
(3) Turn the unit on and make sure the probe selection
LED illuminating corresponds to the connector you
connected probe. If not, press probe selection
button to change it.
Press MODE button to select ARM/LEG pressure
mode. (ARM/LEG LED will illuminate.)
Taking Arm pressure:
(4) Wrap the cuff above the elbow and apply the probe
to the radial or brachial artery at an angle of
approximately 45 to 60 degrees.
(5) Move the probe slowly to locate the point where
the maximum Doppler sounds are heard and the
INF/DUMP LED will blink after a few seconds.
Headset connector
Inflation inlet