18
3-9-3. OPERATION
(1) Connect the probe to the probe connector.
See the step# (1) in the “§.3-1. Turning the unit ON” for the details.
(2) Put ultrasonic gel on the probe top or patient skin.
(3) Turn the unit on and sake sure the probe selection
LED illuminating corresponds to the connector you
connected probe. If not, press probe selection
button to change it.
(4) Put the probe on the middle of the abdomen at
right angle to the skin surface, and move it slowly
to locate the point where the maximum heart beat
Doppler sounds are heard.
Caution:
Verify the fetal heart rate. (Maternal heart sates
match the maternal pulse rates)
(5) When the heart rate becomes stable, press the
probe button or FREEZE button to freeze it.
Skin surface
Gel