Note1: Once you locate the point, please keep the probe there until blood pressure is taken.
Note2: In case INF/DUMP LED doesn't blink due to low flow or arrhythmia, press INF/DUMP
button longer than 2 sec to activate cuff inflator manually in the next step #6.
(6) Press INF/DUMP or probe button to inflate the cuff.
The unit will stop the inflation at estimated 30
mmHg above the sound cessation and deflate the
cuff at 3 to 4 mmHg/sec automatically.
Note: In case the inflation doesn't stop automatically, press
INF/DUMP or Probe button to stop the inflation.
If an error message is displayed in the inflation and
deflation processes, press any button other than probe
selection button to delete the error message and go
back to step #5.
The LCD will show the systolic pressure when the
first Doppler sound is heard and then the cuff
pressure will be dumped automatically.
Press and hold MODE button and press Down
(LEG/TOE) to get on MAGNIFIED waveform mode
if desired. Do it again to go back to FULL waveform
mode.
If the systolic pressure is not satisfied, press and
hold MODE and press Right / Left to move the
cursor (vertical dotted line) to get it adjusted.
(7) Press ARM button to store it as the brachial
pressure and the LCD will show it as shown in the
right.
(8) Press INF/DUMP button to go back to the
measurement mode for ankle pressure.