![Hadeco Smartdop 30EX Скачать руководство пользователя страница 26](http://html.mh-extra.com/html/hadeco/smartdop-30ex/smartdop-30ex_operating-manual_577371026.webp)
19
4. Mode settings
Note: For the mode and menu for PPG probe and PV probe, see “§ 7-3-4. Menu for PPG” and “§ 7-
4-3. Menu for PV”.
4-1-1. Basic Modes
Smartdop has following six Basic Modes:
1. Arterial Blood Velocity:
Mode 1
2. Venous Blood Velocity:
Mode 2
3. Arm and Leg Pressure:
ARM/LEG
4. Toe Pressure:
TOE
5. PV-Arterial:
PV-AC
6. Menu:
For other functions and settings
Pressing MODE button changes basic mode for mode
#1 through 4 and each mode LED selected will
illuminate. (Basic mode can be changed only when on
measurement mode.)
To go to PV-Arterial mode, press Probe Selection
button a few times until both LEDs on the button
illuminate. Press MENU button to go to menu mode
and MENU LED on the button will illuminate.
4-1-2. Measurement / Freeze Mode
Press Freeze button or probe button to freeze or unfreeze the waveform.
Measurement mode:
For measurement (Unfreeze)
Freeze mode:
For observing waveforms and numerical data
Note1: On Freeze mode, following waveforms are temporarily stored in the memory Freeze.
Mode 1:
Waveform of the 5 sec before freezing
Mode 2:
Waveform of the 25 sec before freezing
PV-AC:
Waveform of the 5 sec before freezing
Note2: Pressing probe button also freezes and unfreezes
waveform. Probe button can be set for other function
upon PRB-KEY on menu mode.