3-5. Measuring Blood Velocity
This section explains typical operation
of measuring blood velocity. Refer to Ҥ 5-1. Arterial
and venous blood velocity m
odes” and “§ 4-2. Menu Mode” for mode settings of various
uses.
(1) Connect the probe to the probe connector.
See the step# (1) in the “§.3-1. Turning the unit ON” for the details.
(2) Put ultrasonic gel on tip of probe or patient skin.
(3) Turn the unit on and turn the volume control to
maximum. Make sure the probe selection LED
illuminating corresponds to the connector you
connected probe. If not, press probe selection
button to change it.
(4) Press MODE button to select Mode 1 for arteries
or Mode 2 for veins.
(5) Put the probe on the measurement area and move
it slowly to locate the point where the maximum
Doppler sounds are heard. An ideal probe angle to
the vessel is approximately 45 to 60 degrees.
Wait more than 5 sec for MODE 1 (OR up to 25
sec for MODE 2) without moving probe after the
maximum Doppler sounds, press probe button (or
INF/DUMP button) to freeze the waveform. Press
PRINT button to print the waveform, if necessary.
Note: Function of the probe button can be selected from Freeze, Print, and Freeze & Print on PRB-
Key mode settings. See “§4-2-7-l.OTHERS - PRB-KEY (Probe button function)”.