15
3-7. Toe brachial pressure index (TBI)
Optional PPG probe (PG-21) is required for this testing.
(1) Take the brachial pressure in the same way as ABI. Follow steps #1 to 7 of "§ 3-6
Ankle brachial pressure index (ABI)".
(2) Remove arm cuff from tubing and replace with toe cuff, DVC-1.9 or DPC-2.5 for toe
pressures.
(3) Press MODE button to select toe pressure mode. (TOE LED will illuminate.)
(4) Connect PPG probe to the probe connector # 2 with the round polarity mark up on the
probe connector (12 o'clock).
(5) Press probe selection button to select probe #2.
(6) Wrap the cuff around the base of the big toe and
affix the PPG probe to the pad of the toe with
Velcro strap.
(7) Take the toe pressure the same way for the
brachial. Press the LEG/TOE button to store it as
the toe pressure and the TBI will be displayed as
shown right.
(8) Go to the next toe pressure on the other side.
Note: Press the ARM or LEG/TOE button more than 2 sec to clear the pressure data for the next
patient.