3.2 Thermal Cutoff (instrument only)
If the sample temperature is liable to exceed the compensated temperature range of the
CO
2
or O
2
sensor, you can set an upper temperature limit to automatically cut off the
electrical signal to the sensor to extend the sensor's life.
Under the
MODIFY OPTIONS
menu, highlight
THERMAL CUTOFF
and press
ENTER
.
Choose the sensor (either
CO2
or
O2
).
The CO
2
sensor can be
ENABLED
or
DISABLED
. Select
ENABLED
to move to the data
entry screen, or
DISABLED
to disable this option for the CO
2
sensor.
Once enabled, if the sample temperature exceeds this limit, then alarms are set and a
warning message is sent to the PC. In addition, a
Warning Thermal Cutoff
message
flashes periodically on the instrument LCD.
3.3 Setting Instrument Alarm Limits
If the change product mode is set to
PC ONLY
alarm limits can only be set from the ProBrix Plus PC Program and cannot be changed
directly from the instrument.
It is recommended that you use the ProBrix Plus PC Program to set alarm limits, as the
limits set by the program are stored permanently. If you use the instrument, the values
set here will be overridden by the stored PC values the next time the product is selected.
The instrument provides two alarm relays (
High/Low
and
Low-Low/High-High
), which
can be set separately for CO
2
and Brix/Diet measurement limits. If any measurement is
outside of these limits, an alarm relay is activated as indicated in
* CO
2
, O
2
, or Brix/Diet measurement, as applicable
** On means the alarm relay is closed. Off means the relay is open. Your
Hach Ultra representative can electronically reverse the default open and
closed status.
Selecting the O
2
sensor will take you directly to the data
entry screen.
Enter the thermal cutoff value.
Table 3-4:
Alarm Relay Table
Alarm Condition
High/Low
Relay**
High-High/Low-
Low Relay**
No
alarm: Low Limit < measurement* <High Limit
off
off
Low
alarm: Lo-Low Limit < measurement* < Low Limit
on
off
High
alarm: Hi-High Limit > measurement* > High Limit
on
off
Low-Low
alarm: measurement* < Lo-Low Limit
on
on
High-High
alarm: measurement* > Hi-High Limit
on
on
System
alarm: see
on
on
Содержание ORBISPHERE 3624
Страница 1: ...Operator Manual ORBISPHERE 3624 Revision F 03 10 2008 ...
Страница 2: ......
Страница 14: ...8 of 96 Manual Overview 3624 ProBrix Plus Operator Manual ORBISPHERE ...
Страница 32: ...26 of 96 Installation 3624 ProBrix Plus Operator Manual ORBISPHERE ...
Страница 60: ...54 of 96 PC Program Setup 3624 ProBrix Plus Operator Manual ORBISPHERE ...
Страница 68: ...62 of 96 Calibrations 3624 ProBrix Plus Operator Manual ORBISPHERE ...
Страница 82: ...76 of 96 Accessories and Attachments 3624 ProBrix Plus Operator Manual ORBISPHERE ...
Страница 98: ...92 of 96 Glossary 3624 ProBrix Plus Operator Manual ORBISPHERE ...
Страница 101: ...3624 ProBrix Plus User Notes 95 of 96 Operator Manual ORBISPHERE User Notes ...
Страница 102: ......