GPM-8310 User Manual
90
ESC
This button is used to exit current
screen or return to the main
measurement screen.
Local/ Key Lock
Dual function key. When Remote
mode is activated, press this button to
deactivate Remote mode and switch
to Local mode. When Remote mode is
not activated, this button is used as
lock key of keypad.
Mode
Press the Mode button to select
measurement mode. There are 4
measurement modes.
AC+DC
: Displays all the
components of measurement
signal.
DC
: Displays the DC part of the
measurement signal.
AC
: Displays the AC part of the
measurement signal.
V-MEAN
: Displays the voltage
rectified as a mean value that is
calibrated to RMS value. The value
is same with those obtained from
RMS mode when sine waves are
measured, but it is different when
DC or distorted waves are
measured.