DIGITAL I/O / DA4
125
DA4 Output Function
Overview
You can output voltage, current, active power,
apparent power, reactive power, power factor, phase
angle, frequency, voltage peak, current peak, and
integrated values using a ±5V FS DC voltage.
The output range mode and maximum/minimum
value of manual range mode can only be used when
using a remote control interface. Likewise this
parameter can only be configured via remote control.
Please see the commands on page 157 for full usage
details.
Output Format
You can select a preconfigured output format or
configure your own original format.
Preconfigured
Format
Normal
Integrator
Ch1
V
P
Ch2
I
WP
Ch3
P
q
Ch4
VHz
VHz
Rated Integration
Time
In the D/A output of integrated values, 5.0 V FS
represents the integrated value when the rated range
value is applied for the rated integration time. The
default setting is 1.00.00 (1 h, 0 min, 0 s).
If you set the rated integration time to 0.00.00, the
D/A output value will be 0 V.
Output Range
Mode
The DA4 have two output range mode: Fixed range
mode or Manual range mode. The default setting is
Fixed.
Fixed (Fixed
range mode)
When a measurement function’s rated value is
received, +5V is output.
Содержание GPM-8310
Страница 1: ...Digital Power Meter GPM 8310 USER MANUAL Rev A ISO 9001 CERTIFIED MANUFACTURER ...
Страница 74: ...GPM 8310 User Manual 74 3 Press Config soft key to Enter SYSTEM CONFIG setting screen ...
Страница 135: ...DIGITAL I O DA4 135 Pin Diagram Use the built in power supply Use the external power Method of series ...
Страница 221: ...COMMAND OVERVIEW 221 Example STATUS QMESSAGE ON STATUS QMESSAGE STATUS QMESSAGE 1 ...
Страница 239: ...APPENDIX 239 Status system The diagram below is a description of the status system ...
Страница 243: ...APPENDIX 243 Dimensions Units mm ...