APPENDIX
233
Accuracy of apparent
power S
voltage ac current accuracy
Accuracy of reactive power
Q
accuracy of apparent power + (
√
1.0004 -
λ
2) - (
√
1 -
λ
2) ×100 %
Accuracy of power factor
λ
± [(
λ
-
λ
/1.0002)+
丨
cosø-cos{ø+sin-1 (influence from
the power factor when
λ
= 0%/100)}
丨
] ±1 digit
when voltage and current are at the measurement
range rated input
Accuracy of phase
difference
Φ
± [
丨
ø-cos-1(
λ
/1.0002)
丨
+ sin-1 (influence from the
power factor when
λ
= 0 % / 100)] ± 1 digit when
voltage and current are at the measurement range
rated input
Accuracy when the crest
factor is set to 6 or 6A
accuracy obtained by doubling the measurement
range error for the accuracy when the crest factor is
set to 3
Accuracy changes caused
by data update interval
When the data update interval is 100 ms, and Auto,
add 0.05% of reading to the 0.1 Hz to 1 kHz
accuracy.
* f is the frequency of input signal in kHz
Voltage, Current and Active Power Measurements
Item
Specifications
Measurement method
Digital sampling method
Crest factor
3 or 6 (6A)
Wiring system
Single-phase , two-wire(1 P2 W)
Range select
Select manual or auto ranging
Auto range
Auto-range increase
The range is upped when any of the following
conditions is met.
Crest factor 3 Urms or Irms exceeds 130% of the
currently set measurement range.
Upk, Ipk value of the input signal
exceeds 300% of the currently set
measurement range.
Crest factor 6 Urms or Irms exceeds 130% of the
currently set measurement range.
Upk, Ipk value of the input signal
exceeds 600% of the currently set
measurement range.
Crest factor 6A Urms or Irms exceeds 260% of the
currently set measurement range.
Upk, Ipk value of the input signal
exceeds 600% of the currently set
measurement range.
Содержание GPM-8310
Страница 1: ...Digital Power Meter GPM 8310 USER MANUAL Rev A ISO 9001 CERTIFIED MANUFACTURER ...
Страница 74: ...GPM 8310 User Manual 74 3 Press Config soft key to Enter SYSTEM CONFIG setting screen ...
Страница 135: ...DIGITAL I O DA4 135 Pin Diagram Use the built in power supply Use the external power Method of series ...
Страница 221: ...COMMAND OVERVIEW 221 Example STATUS QMESSAGE ON STATUS QMESSAGE STATUS QMESSAGE 1 ...
Страница 239: ...APPENDIX 239 Status system The diagram below is a description of the status system ...
Страница 243: ...APPENDIX 243 Dimensions Units mm ...