BASIC SETTING
37
Setting up auto-zero function
Steps
1.
Press
Setup
button.
2.
Press
Enter
button.
3.
Press down arrow key to move
cursor to the
Auto Zero
field.
4.
Use soft keys to select and confirm
the desired option.
Option
On Auto-zero function is activated once per
hour or when range is switched.
Off
Auto-zero function is only activated once
when the range is switched. The auto-zero
function is turned off when the integrator
function is executed.
Default value
Off