APPENDIX
235
Frequency
IHz and VHz
Voltage Peak
V+pk and V-pk
Current Peak
I+pk and I-pk
Active Power Peak
P+pk and P-pk
Total Harmonic Distortion
THDI and THDV
Mathematical Computation MATH
Maximum Current Ratio
MCR
Frequency Measurement
Item
Specifications
Measurement item
Voltage and current
Measurement frequency
range
Data update interval Measurement Frequency Range
0.1 s
20 Hz
≤
f
≤
100 kHz
0.25 s
10 Hz
≤
f
≤
100 kHz
0.5 s
5 Hz
≤
f
≤
100 kHz
1 s
2.0 Hz
≤
f
≤
100 kHz
2 s
1.0 Hz
≤
f
≤
100 kHz
5 s
0.5 Hz
≤
f
≤
100 kHz
10 s
0.2 Hz
≤
f
≤
100 kHz
20 s
0.1 Hz
≤
f
≤
100 kHz
Auto ( * )
0.1 Hz
≤
f
≤
100 kHz
( * ) Limit of the measurement lower limit frequency
by the Timeout setting
Timeout
lower limit frequency
1 s
2.0 Hz
5 s
0.5 Hz
10 s
0.2 Hz
20 s
0.1 Hz
Measurement range
Auto switching among six types: 100mHz, 1 Hz, 10
Hz, 100 Hz, 1 kHz, 10 kHz, and 100 kHz.
Frequency filter
Select OFF or ON (cut off frequency of 500 Hz)
Accuracy
Requirements
When the input signal level is
30% or more of the
measurement range If the
crest factor is set to 3.
(60% or more if the crest
factor is set to 6 or 6A)
• Frequency filter is ON when
measuring voltage or current
of 200 Hz or less.
± (0.06% of reading)
Содержание GPM-8310
Страница 1: ...Digital Power Meter GPM 8310 USER MANUAL Rev A ISO 9001 CERTIFIED MANUFACTURER ...
Страница 74: ...GPM 8310 User Manual 74 3 Press Config soft key to Enter SYSTEM CONFIG setting screen ...
Страница 135: ...DIGITAL I O DA4 135 Pin Diagram Use the built in power supply Use the external power Method of series ...
Страница 221: ...COMMAND OVERVIEW 221 Example STATUS QMESSAGE ON STATUS QMESSAGE STATUS QMESSAGE 1 ...
Страница 239: ...APPENDIX 239 Status system The diagram below is a description of the status system ...
Страница 243: ...APPENDIX 243 Dimensions Units mm ...