Query Syntax
:RATE:AUTO:TIMeout?
Parameter/
Return parameter
<TIME>
1, 5, 10, 20(s)
Example
RATE:AUTO:TIMEOUT 1
:RATE:AUTO:TIMEOUT?
-> :RATE:AUTO:TIMEOUT 1
:RATE:AUTO:SYNChronize
Description
Sets or returns the synchronization source for
when the data update interval is set to Auto.
Syntax
:RATE:AUTO:SYNChronize {U1|I1}
Query Syntax
:RATE:AUTO:SYNChronize?
Parameter/
Return parameter
U1
Select synchronize source at voltage.
I1
Select synchronize source at current.
Example
:RATE:AUTO:SYNCHRONIZE U1
:RATE:AUTO:SYNCHRONIZE?
-> :RATE:AUTO:SYNCHRONIZE U1
Set
Query