HANDLER/SCAN INTERFACE
85
the measured value from the reference value as a
percentage.
{ [(Measured Value-Reference)/Reference]%}.
A measured value that falls within the upper and lower
limits is considered IN (pass), a value that falls below the
lower limits is considered LO, and a value that falls over
the upper limit is a HI.
Reference
value
Upper
limit
Lower
limit
IN
HI
LO
For both scan modes, the IN, HI or LO will be shown on
the display for each judgment (if the time between each
judgment is not too fast).
Display Overview
Scan function
indicator
Ready to start scan message
Reference, limits, scan mode, current
channel, measurement delay
Change
display view
1. Select the
Scan function
Press
Scan
Scan to access the scan mode, as shown
above.
2. Select the
compare mode
Use the arrow keys to navigate to the Mode setting. Press
the Enter key to toggle the compare mode.
Mode
Move
Toggle
Enter
Range
Abs,
△
%