Binning Function
Background
The Binning function is used to grade DUTs into eight
different bins according to 8 sets of upper and lower
limits. Two compare modes can be used in this function,
ABS and
△
% modes.
Binning function
indicator
Upper and lower
limits for the 8
bins
Grading results
Reference, compare mode, beep
mode and display mode
1. Select the
Binning function
Press the
Binning
key to access this function.
2. Select the
compare mode
Use the arrow keys to go to the Mode setting.
Press Enter to toggle between ABS or
△
% compare
modes.
Mode setting
Move
Toggle
Enter
ABS Mode
The ABS mode allows you to set
the upper and lower limits of each
bin as absolute resistance values.
△
%
The Delta % mode allows you to
set the upper and lower limits of
each bin as percentage value from
the reference value.