MEASUREMENT
61
Measure Delay
Background
The Measure Delay setting inserts a delay time between
each measurement. Measure delay is turned off by
default.
Test signal
Default Measurement
start time
Measurement start with
Measure delay time
Measure delay time
The measure delay setting is useful for measuring
components that need some time to charge if the default
measurement start time is not adequate. An adequate
delay time allows the meter to avoid the effects of
transient disturbances that are usually seen when
measuring reactive DUTs with a current source.
1. Select
Measure Delay
setting
From one of the main screens, press
the
ESC
key so that the menu
system at the bottom of the display
has focus.
Go to Meas. Setup and press Enter.
Go to Measure Delay and press
Enter.
Meas. Setup
menu icon
Move
Select menu
or setting
Enter
2. Measure Delay
setting appears
Use the arrow keys to turn Measure Delay on and set the
delay time. Press Enter to confirm the setting.
Measure delay
setting
Measure Delay* OFF, ON: 000.000 ~ 100.000s
* When the set value is > 0.1s, the resolution is 0.1s.
When the set value is < 0.1S, the resolution is 1mS.