GOM-804 & GOM-805 User Manual
110
Example
BINN1:LIM:UPP 0.95,maohm
Sets bin1 upper limit value to 0.95M
Ω
.
BINN1:LIM:UPP?
>0.9500E+6
Returns the upper limt as 0.95M
Ω
.
BINNing<X>:PERCent:LOWer
Set
Query
Description
Sets or returns the lower value percentage value for the
selected bin. The value is a percentage offset from the
reference value.
Syntax
Query Syntax
BINNing<X>:PERCent:LOWer <NRf>
BINNing<X>:PERCent:LOWer?
Parameter
<X>
1~8
<NRf>
000.00~999.99
Return parameter <NR2>
000.00~999.99
Example
BINN1:PERC:LOW 10.15
Sets the bin1 lower limit percent value to -10.15%.
BINN1: PERC:LOW?
>10.15
Returns the lower limit percentage value as -10.15%.
BINNing<X>:PERCent:UPPer
Set
Query
Description
Sets or returns the upper value percentage value for the
selected bin. The value is a percentage offset from the
reference value.
Syntax
Query Syntax
BINNing<X>:PERCent:UPPer <NRf>
BINNing<X>:PERCent:UPPer?
Parameter
<X>
1~8
<NRf>
000.00~999.99
Return parameter <NR2>
000.00~999.99
Example
BINN1:PERC:UPP 150.95
Sets the bin1 upper limit percent value to +150.95%.
BINN1:LIM:UPP?
>150.95
Returns the upper limit percentage value as +150.95%.
www.
GlobalTestSupply
.com
Find Quality Products Online at: