GOM-804 & GOM-805 User Manual
118
DPER
The test results are judged from a
reference value ± a percentage offset.
(delta percent)
Example
CALC:SCAN:LIM:MODE ABS
Sets compare mode to absolute values.
CALCulate:SCAN:LIMit:REFerence
Set
Query
Description
Sets or returns the reference limit for the scan function.
Syntax
Query Syntax
CALCulate:SCAN:LIMit:REFerence {<NRf>[,<String>]}
CALCulate:SCAN:LIMit:REFerence?
Parameter
<NRf>
000.0001~999.9999
<String>
mohm/ohm/kohm/maohm,unit
If unit is not set,the unit will be
automatically set by the present range.
Return parameter <NR3>
000.0001~999.9999E± X
Example
CALC:SCAN:LIM:REF 10.00,mohm
Sets the reference limit to 10.00m
Ω
.
CALC:SCAN:LIM:REF?
>10.0000E-3
Returns the reference limit as 10.00m
Ω
.
CALCulate:SCAN:LIMit:UPPer
Set
Query
Description
Sets or returns upper limit of the scan function.
Syntax
Query Syntax
CALCulate:SCAN:LIMit:UPPer {<NRf>[,<String>]}
CALCulate:SCAN:LIMit:UPPer?
Parameter
<NRf>
000.0000~999.9999
<String>
mohm/ohm/kohm/maohm,unit
If unit is not set, the unit will be
automatically set by the present range.
Return parameter <NR3>
000.0000~999.9999E± X
Example
CALC:SCAN:LIM:UPP 1.37,kohm
Sets the upper limit to 1.37k
Ω
.
CALC:SCAN:LIM:UPP?
>1.3700E+3
Returns the upper limit as 1.37k
Ω
.