GOM-804 & GOM-805 User Manual
74
*The OFF operator sets the Logic as
true when Operand1 is true.
** Bin O is defined as outside bin 1~ 8.
Note
The Bin logic settings are not available for the GOM-804.
Pressing ESC before pressing ENTER will exit from the
selected External I/O setting.
Handler Mode
Background
The Handler Mode setting determines the behavior of
the result signals from the handler interface. There are
two settings, Clear and Hold. The Clear setting will clear
the results of the previous test before starting the
succeeding one and the Hold setting will keep the test
result of the previous test until the succeeding test has
completed.
The timing diagrams below are used as examples. All the
result signals in the examples are active high.
Clear example
Clear: All result signals (PASS, Fail, High and Low) are
cleared at the falling edge of EOT and the results from
the current test are output at the rising edge of the EOT
signal.
Previous
results
cleared
New
results
Trigger
Ready
EOT
Pass
Fail
High
Low
EOT falling
edge
EOT rising
edge
Previous
results