GETTING STARTED
17
Speed
The Speed key toggles between 10
samples per second and 60 samples per
second (Slow rate and Fast rate).
REL
The REL key is used to perform a zero
adjustment to the test leads or a DUT.
RT
The RT key is used to display the
real-time (not averaged) measured
resistance value.
Scan
The Scan key is used to turn on the
Scan function.
Dry
The Dry key is used to turn on the dry
circuit measurement mode which allows
the GOM-805 to measure the contact
resistance of switches and connectors
according to DIN IEC 512 and ASTM
B539 standards. GOM-805 only.
Trigger
When in the internal trigger mode,
pressing the Trigger key will turn on the
external trigger mode. When in the
external trigger mode, pressing the
Trigger key will perform a manual
trigger.
A long press of the Trigger key when in
external trigger mode will reset the
trigger mode back to the internal trigger
mode.
Display
The Display key toggles between the
standard display mode and the
simplified display mode (sans menus
and display icons).
Local
The LOCAL
key will switch the
milliohm meter between local and
remote mode.
Diode
The Diode key is used to turn on the
Diode measurement function.
www.
GlobalTestSupply
.com
Find Quality Products Online at: