Immunity test
IEC 60601 test level
Compliance level
E l e c t r o m a g n e t i c
environment - guidance
Conducted RF lEC
61000-4-6
Radiated RF lEC
61000-4-3
3 Vrms
150 kHz to 80 MHz
6 Vrms in ISM bands
3 V/m
80 MHz to 2.7 GHz
3 8 5 M H z - 5 7 8 5 M H z
Test specifications for
E N C L O S U R E P O R T
I M M U N I T Y t o R F
wireless communication
equipment
(Refer to table 9 of IEC
60601-1-2:2014)
3 Vrms
150 kHz to 80 MHz
6 Vrms in ISM bands
3 V/m
80 MHz to 2.7 GHz
3 8 5 M H z - 5 7 8 5 M H z
Test specification s for
E N C L O S U R E P O R T
IMMUNITY to RF wireless
communicati on equipment
(Refer to table 9 of IEC
60601-1-2:2014)
P o r t a b l e a n d m o b i l e
R F c o m m u n i c a t i o n s
e q u i p m e n t s h o u l d b e
used no closer to any part
of the models Minipex,
including cables, than the
recommended separation
distance calculated from
the equation applicable
to the frequency of the
transmitter.
Recommended separation
distance
d=[3,5/V1]×P
1/2
d=1.2×P
1/2
80 MHz to 800
MHz
d=2.3×P
1/2
800 MHz to 2.5
GHz