6 Cleaning, Disinfection and Sterilization
The cleaning, disinfection and sterilization of lip hook, file clip and touch probe are
as follow.
Unless otherwise stated, they will be hereinafter referred to as “products”.
Warnings:
The use of strong detergent and disinfectant (alkaline pH>9 or acid pH <5) will
reduce the life span of products. And in such cases, the manufacturer takes no
responsibility.
This device shall not be exposed to high temperature above 138℃.
Processing limit
The products have been designed for a large number of sterilization cycles. The
materials used in manufacture were selected accordingly. However with every
renewed preparation for use, thermal and chemical stresses will result in ageing of
the products. The maximum number of sterilizations for file clip is 200 times. For lip
hook and touch probe, it is 1000 times.
6.1 Initial processing
6.1.1 Processing principles
It is only possible to carry out effective sterilization after the completion of effective
cleaning and disinfection. Please ensure that, as part of your responsibility for the
sterility of products during use, only sufficiently validated equipment and product-
specific procedures are used for cleaning/disinfection and sterilization, and that the