![Guilin Woodpecker Medical Instrument Minipex Скачать руководство пользователя страница 28](http://html.mh-extra.com/html/guilin-woodpecker-medical-instrument/minipex/minipex_instruction-manual_2278509028.webp)
26
with the European Pharmacopoeia or the United States Pharmacopoeia).
f) After cleaning, the chemical residue should be less than 10mg / L.
g) The air used for drying must be filtered by HEPA.
h) Regularly repair and inspect the disinfector.
6.5 Drying
If your cleaning and disinfection process does not have an automatic drying
function, dry it after cleaning and disinfection.
Methods:
1.Spread a clean white paper (white cloth) on the flat table, point the product against
the white paper (white cloth), and then dry the product with filtered dry compressed
air (maximum pressure 3 bar). Until no liquid is sprayed onto the white paper (white
cloth), the product drying is completed.
2.It can also be dried directly in a medical drying cabinet (or oven). The
recommended drying temperature is 80℃~120℃ and the time should be 15~40
minutes.
Note:
a)The drying of product must be performed in a clean place.
b)The drying temperature should not exceed 138 °C;
c)The equipment used should be inspected and maintained regularly.
6.6 Inspection and maintenance
1.Check the product. If there is still visible stain on the product after cleaning/