Problems
Possible causes
Solutions
No power and no signal on
the screen after the power on.
1. If the battery is placed correctly?
2. If the battery with no power?
1. Re-install the battery.
2. Recharge the battery.
The length of the root canal
cannot be measured.
1. If the measuring wire is connected
correctly?
2. If the measuring wire is broken?
Confirm the measuring wire is plugged
firmly, link the lip hook with the file clip
to check if the measuring wire is broken.
No sound of alarm.
If the volume is set at “mute”?
Adjust the sound level.
The “charging prompt”
icon didn’t blinking when
charging battery.
1.The adapter is not connected well.
2.Have used faulty adapter with excessive
output.
3.The battery is not installed well.
4.The battery has been damaged.
1.Reconnect the adapter.
2.Change the adapter, must use the
original adapter.
3.Reinsert the battery and then reconnect
the adapter.
4.Change the battery and then reconnect
the adapter.
Display not steady while
measuring: the measurement
result is rather longer or
shorter; numerical display
irregular.
If the connection between the lip hook
and the oral mucosa is ok?
Make sure the lip hook has contacted the
oral mucosa at a good position.
Is there a blood/saliva overflowing, glued
to the crown?
Blood, liquid overflow from the root
canal, glued to the crown or the tooth
neck, will cause short-circuit then cause
the in-normal phenomena. Clean the blood
and the liquid.