29
The procedure used here is to perform steam sterilization through three pre-vacuums.
6.9 Storage
1. Store in a clean, dry, ventilated, non-corrosive atmosphere with a relative humidity
of 10% to 93%, an atmospheric pressure of 70 kPa to 106KPa, and a temperature of
-20 °C to +55 °C;
2. After sterilization, the product should be packaged in a medical sterilization bag
and stored in a special storage cabinet. The storage time should not exceed 7 days. If
it is exceeded, it should be reprocessed before use.
Notes:
a)The storage environment should be clean and must be disinfected regularly;
b)Product storage must be batched and marked and recorded.
6.10 Transportation
1. Prevent excessive shock and vibration during transportation, and handle with care;
2. It should not be mixed with dangerous goods during transportation.
3. Avoid exposure to sun or rain or snow during transportation.
The cleaning and disinfection of main unit and the measuring wire are as follows.
• Before each use, wipe the surface of the machine and the measuring wire with a
soft cloth or paper towel soaked in 75% medical alcohol. Repeat the wipe for at least
3 times.