Surge
lEC 61000-4-5
±1 kV line to line
±2 kV line to earth
±1 kV line to line
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11.
<5 % UT
(>95% dip in UT.)
for 0.5 cycle
40 % UT
(60% dip in UT) for
5 cycles
70% UT
(30% dip in UT) for
25 cycles
<5% UT
(>95 % dip in UT)
for 5 sec
<5 % UT
(>95% dip in UT.)
for 0.5 cycle
40 % UT
(60% dip in UT) for 5
cycles
70% UT
(30% dip in UT) for 25
cycles
<5% UT
(>95 % dip in UT)
for 5 sec
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the models
Minipex require continued operation
during power mains interruptions, it is
recommended that the models Minipex
be powered from an uninterruptible
power supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
lEC 61000-4-8
30A/m
30A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
NOTE
U
T
is the a.c. mains voltage prior to application of the test level.
Guidance & Declaration - Electromagnetic immunity
The models Minipex are intended for use in the electromagnetic environment specified below. The
customer or the user of the models Minipex should assure that it is used in such an environment.