25
Note:
Before use, you must carefully read the operating instructions provided by the
equipment manufacturer to familiarize yourself with the disinfection process and
precautions.
b) With this equipment, cleaning, disinfection and drying will be carried out together.
c) Cleaning: (c1) The cleaning procedure should be suitable for the product to be
treated. The flushing period should be sufficient (5-10 minutes). Pre-wash for 3
minutes, wash for another 5 minutes, and rinse it for twice with each rinse lasting for
1 minute. (c2) In the washing stage, the water temperature should not exceed 45 °C,
otherwise the protein will solidify and it is difficult to remove. (c3) The solution used
can be pure water, distilled water, deionized water or multi-enzyme solution, etc.,
and only freshly prepared solutions can be used. (c4)During the use of cleaner, the
concentration and time provided by manufacturer shall be obeyed. The used cleaner
is neodisher MediZym (Dr. Weigert).
d) Disinfection: (d1) Direct use after disinfection: temperature ≥ 90 ° C, time ≥ 5 min
or A0 ≥ 3000;
Sterilize it after disinfection and use: temperature ≥ 90 ° C, time ≥ 1 min or A0 ≥ 600
(d2) For the disinfection here, the temperature is 93 ° C, the time is 2.5 min, and
A0>3000
e) Only distilled or deionized water with a small amount of microorganisms (<10 cfu/
ml) can be used for all rinsing steps. (For example, pure water that is in accordance