Emissions test
Complianc
Electromagnetic environment - guidance
RF emissions CISPR 11
Group 1
The models Minipex use RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions CISPR11
Class B
The models Minipex are suitable for used in
domestic establishment and in establishment
directly connected to a low voltage power supply
network which supplies buildings used for
domestic purposes.
Harmonic emissions lEC 61000-3-2
Class A
Voltage fluctuations / flicker emissions
lEC 61000-3-3
Complies
Guidance & Declaration — electromagnetic immunity
The models Minipex are intended for use in the electromagnetic environment specified below. The
customer or the user of the models Minipex should assure that It is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
lEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV,
±15 kV air
±8 kV contact ±2 kV, ±4
kV, ±8 kV, ±15 kV air
Floors should be wood, concrete
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
E l e c t r i c a l f a s t
transient/burst
IEC 61000-4-4
± 2 k V f o r p o w e r
supply lines
±1 kV for Input/
output lines
± 2 k V f o r p o w e r
supply lines ±1kV for
interconnecting cable
Mains power quality should be that
of a typical commercial or hospital
environment.